Troubleshooting Y Motor Errors on the MiSeq
Last updated
Last updated
© 2023 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. Trademark information: illumina.com/company/legal.html. Privacy policy: illumina.com/company/legal/privacy.html
The following MiSeq Y motor errors usually stop the run and indicate communication problems with the stage or problems with stage hardware:
Y Motor: Closed-loop move cannot reach target position: Start = 75.4150390625 mm, Target = 75.40472 mm, Distance = -0.0103190625000025 mm, Stopped at 74.42236328125 mm, ExceptionMessage =
The FPGA reported an error while executing command "YMVTO 154429"; error message: "Error: too many steps needed"
Error: Timeout of 60.000 seconds exceeded while waiting for FPGA command "YMVTILE 1327" to complete
Troubleshooting Steps:
Perform the post-run wash, if the run stops or after the run is completed.
Power cycle\* the MiSeq.
If no issues with re-initialization, perform the Y Stage test 5 times.
Make sure that a used flow cell is loaded on the instrument (the same wash flow cell should be okay).
Select Y Stage test.
MiSeq Control Software (MCS) v 4.0: From Main Menu, select System Check.
MCS 3.1 and older: From Welcome Screen, select Manage Instrument, select System Check.
Follow the prompts on the screen to start the test, and repeat the test 4 more times (5 times total).
If all Y Stage tests pass continue with normal sequencing.
If the system fails to initialize or any of the 5 replicates of the Y Stage test fail, contact Illumina Technical Support for further assistance.
For further information, search for the following Knowledge Base articles:
\*How to Shut down or power cycle the MiSeq.
*\How To Power Cycle the MiSeq Video.
System Checks on MiSeq.
For any feedback or questions regarding this article (Illumina Knowledge Article #1452), contact Illumina Technical Support techsupport@illumina.com. |