> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000001367.md).

# System error detected please power cycle and perform system checks error on the iSeq 100

The "System error detected please power cycle and perform system checks error" on the iSeq 100 can be caused by multiple factors and indicates that the instrument experienced some type of communication error, usually caused by one of the following.

#### 1. Sample sheet in manual mode saved to network or thumb drive.

* The error will likely occur during a sequencing run causing it to abort prematurely.
* Mitigation: Always save the sample sheet locally on the instrument.

#### 2. Instrument is not connected to the internet, and the control software is looking for software updates.

* Mitigation: Do one of the following.
  * Connect the instrument to the internet.
  * Update settings to turn off "Autocheck for software updates" in the "System Customization" tab of "System Settings."

3. Incorrectly Named Drives - If the "Data" drive is set to a letter drive other than "D:", this can cause this error to occur

* Check the D:\ drive name convention.
* If a different drive is named "Data", Contact Illumina Technical Support

#### 4. Network issues

* To confirm, restore to factory default settings and **do not** reconnect to the network once completed. Perform runs locally to see if the error repeats. If the error does not repeat, the error is likely caused by the network setting. For further information, search for Knowledge Base article **Restore to Factory Settings Option on iSeq 100**.

If the issue persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1367), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001367%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000001367.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
