Xp reagents components and stability on the NovaSeq 6000
Last updated
Last updated
© 2023 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. Trademark information: illumina.com/company/legal.html. Privacy policy: illumina.com/company/legal/privacy.html
NovaSeq 6000 Xp kits come in 2 configurations:
2 lane kits are for use with the S1 and S2 reagent kits.
4 lane kits are for use with S4 kits.
Each kit contains ExAmp reagents and a 2- or 4-lane manifold.
Note: the reusable flow cell dock is sold separately.
Flow Cell Dock
Xp flow cell dock is reusable.
Flow cell is inverted and manifold is fitted over the flow cell.
Xp Kit components, shipping conditions, and storage temperatures:
Quantity | Kit Component | Shipping Condition | Storage Temp |
---|---|---|---|
1 | DPX1/JPX1 | Dry ice | -20°C |
1 | DPX2/JPX2 | Dry ice | -20°C |
1 | DPX3 | Dry ice | -20°C |
1 | Xp Manifold | Dry ice | \*Can leave frozen with the kit or store at room temperature (RT) |
Thaw components at RT for 10 min then set aside on ice.
Mix and use within 30 minutes at room temperature or 1 hour if placed on ice.
If unmixed with library pool, Xp reagents can be refrozen once (freeze as quickly as possible after thawing).
Extra mastermix should be discarded.
Wells on the manifold are numbered right to left to match the lane number of an inverted flow cell.
Magnetic clamp holds in place when on dock.
Workflow
Start the run within 30 minutes after loading library pools on the flow cell.
Note: prolonged incubation might reduce the percentage of clusters passing filter (%PF).
For any feedback or questions regarding this article (Illumina Knowledge Article #2422), contact Illumina Technical Support .