> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002422.md).

# Xp reagents components and stability on the NovaSeq 6000

NovaSeq 6000 Xp kits come in 2 configurations:

* **2 lane kits are for use with the S1 and S2 reagent kits.**
* **4 lane kits are for use with S4 kits.**

Each kit contains ExAmp reagents and a 2- or 4-lane manifold.

* **Note**: the reusable flow cell dock is sold separately.

**Flow Cell Dock**\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-297f13fd4840c093eb89636c555fbbab6396cfbc%2Fimage1.png?alt=media\&token=cf766355-02c8-4fba-8c17-7595dcd92dde)

* Xp flow cell dock is reusable.
* Flow cell is inverted and manifold is fitted over the flow cell.

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-dc9afacb148b2c28b797330201b54b60bafb5567%2Fimage2.png?alt=media\&token=b99c0d90-5b3b-4407-884c-c72751df02ab)

**Xp Kit components, shipping conditions, and storage temperatures:**

| Quantity | Kit Component | Shipping Condition | Storage Temp                                                      |
| -------- | ------------- | ------------------ | ----------------------------------------------------------------- |
| 1        | DPX1/JPX1     | Dry ice            | -20°C                                                             |
| 1        | DPX2/JPX2     | Dry ice            | -20°C                                                             |
| 1        | DPX3          | Dry ice            | -20°C                                                             |
| 1        | Xp Manifold   | Dry ice            | \*Can leave frozen with the kit or store at room temperature (RT) |

* Thaw components at RT for 10 min then set aside on ice.
* Mix and use within 30 minutes at room temperature or 1 hour if placed on ice.
* If unmixed with library pool, Xp reagents can be refrozen once (freeze as quickly as possible after thawing).
* Extra mastermix should be discarded.
* Wells on the manifold are numbered right to left to match the lane number of an inverted flow cell.
* Magnetic clamp holds in place when on dock.

**Workflow**

* Start the run within 30 minutes after loading library pools on the flow cell.
  * **Note**: prolonged incubation might reduce the percentage of clusters passing filter (%PF).

More information, please reference [NovaSeq 6000 Sequencing System Guide](https://support.illumina.com/sequencing/sequencing_instruments/novaseq-6000/documentation.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2422), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002422%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002422.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.