> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001252.md).

# Best practices for manually normalizing library concentrations

Library normalization is the process of diluting libraries of variable concentration to the same concentration before volumetric pooling, ensuring an even read distribution for all samples. Normalization best practices can be used for any Illumina library preparation requiring a manual normalization. Steps for normalization are:

1. Determine the library size.
2. Quantify the libraries.
3. Plan the dilution calculations.
4. Pool the normalized libraries

**Note**: Some library preparation methods, such as Nextera XT and Illumina DNA Prep, (M) Tagmentation, include internal normalization strategies. Manual normalization is not needed when another normalization method is used. However, when the input amount is lower than 100 ng for Illumina DNA Prep, (M) Tagmentation or the final library yield is less than 10-15 nM for Nextera XT, the internal normalization is not compatible. In these situations, it is necessary to perform manual normalization

* The normalization mechanisms are different in the Illumina DNA Prep, (M) Tagmentation and Nextera XT workflows; because of this difference, the library denature and dilution procedures prior to sequencing are different.

**Determine the library size**

Validate the library size by running the samples on a Bioanalyzer, Fragment Analyzer, or equivalent. This step also provides visibility into possible library issues, such as adapter dimers or unexpected library sizes.

**Quantify the libraries**

Quantify the libraries using the recommended quantification method as specified in the library preparation guide, summarized in the [Library quantification and quality control quick reference guide](/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001247.md).

Convert this value from ng/µl to nM using the average library size obtained from the Bioanalyzer or Fragment Analyzer, and the [Converting ng/µl to nM when calculating dsDNA library concentration](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/LP-Converting-ng-µl-to-nM-when-calculating-dsDNA-library-concentration/README.md).

**Plan the dilution calculations**

The concentration normalization occurs in this step. Dilution can be done using molecular grade water or 10 mM Tris-HCl pH 8.5. The [Pooling calculator](https://support.illumina.com/help/pooling-calculator/pooling-calculator.htm) can also be used to calculate library dilutions, and can be used if the libraries have the same or different concentration.

1. Determine the common concentration to dilute the libraries for subsequent applications. For most Illumina sequencing platforms, 2-4 nM for each library is the preferred starting concentration for the denaturation and dilution guidelines; consult the respective instrument user guides for more information.
2. Calculate the dilution of the libraries using the following equation.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a86c463d1e8132812e9b61c366f311ffbaed2f4b%2Fimage1.png?alt=media\&token=e92b18a1-329a-4943-82bc-5dfe2af4e9ce)

For the most even representation of samples and most reliable cluster density, make sure that all pipetted volumes are at least 2 µl, though a minimum volume of 5 µl is ideal. Pipetting very low volumes can introduce significant concentration errors. Adjust calculations to make sure that appropriate volumes are used. For highly concentrated libraries, use one or more intermediate dilutions.

For example, the following calculations demonstrate how to dilute 3 libraries (with different starting concentrations of 15 nM, 20 nM, and 50 nM) to a final concentration of 4 nM in a final volume of 15 µl for each library.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-fce1efc723a9169742f5ea23e4a89974da151123%2Fimage2.jpg?alt=media\&token=919d413f-b3f5-4179-bdb3-c533369af7c1)

\* The 50 nM library example only uses 1.2 µl of stock DNA to dilute down to 4 nM library. For the most accurate results, an intermediate dilution of 20 nM is necessary to achieve at least a 2 µl pipette volume.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6cdc23b1cd4f802940c61da6fe0157cc96f109a5%2Fimage3.jpg?alt=media\&token=446611b2-cff1-4322-b166-b3a2e8dec83e)

3. Dilute the libraries according to the calculations above. The libraries are now normalized.

**Pool the normalized libraries**

Volumetric pooling: Combine equal volumes of each normalized library into a microcentrifuge tube and gently pipette contents up and down 10 times to mix thoroughly.

**Next steps**

The normalized pool is now ready to be denatured and sequenced. For detailed instructions on denaturation procedures, use the appropriate instrument denaturation and dilution guide.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1252), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001252%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001252.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
