> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001507.md).

# Proper care of Tecan/IAPS with IAC robot fluidics

This article outlines the recommended care and maintenance procedures for the Infinium Automated Pipetting System (IAPS) used with Illumina Automation Control (IAC) fluidics. Proper maintenance is essential to ensure the robot operates reliably and maintains accurate and consistent pipetting performance throughout its use.

**System Liquid**\
Ensure that the system liquid carboy contains an adequate volume of clean system liquid (de-ionized water).

* Always confirm that the intake tubing is below the system liquid line in the carboy.
* Fill the carboy frequently to prevent the risk of the fluid level running too low.
* Keep the carboy opening covered to prevent contamination from entering the system liquid.
* Periodically empty and clean the carboy as the interior surfaces can become moldy over time.

**Priming**

Make sure that the fluidics lines are fully primed before use. Fully primed lines are especially important after periods of inactivity.

* Conduct a tip bleach wash followed by a minimum of 3 Sys Wash routines before use. Additional Sys Wash routines can be run as desired.
* Watch for excessive bubbles in the syringes during pipetting steps. If bubbles are present, repeat additional Sys Wash routines.
* If the system liquid carboy runs dry, conduct repeated Sys Wash routines to prime the fluidics lines after refilling the carboy.

**Bleach Washing**

Perform periodic bleach washes to minimize potential growth in fluidics lines and to clean the tips.

* Perform a tip bleach wash (Robot QC Tasks > Robot Bleach Wash > Option 0) daily before each use. Immediately remove the bleach trough from the robot deck after use to help eliminate bleach fumes.
* Do not perform a system bleach wash.

**Water Circulator for Te-Flow**

Check for proper fluid levels in the water circulator, which controls the temperature of the Te-Flow unit.

* Before each use, open the lid and make sure that water is above the heating coils. If the fluid level is low, replenish as necessary with distilled water prepared with anti-microbial additive. It is not necessary to drain the existing fluid.
* Confirm that the lid is securely in place to minimize evaporation.
* Refer to the Infinium Assay documentation for details on setting up the water circulator and Te-Flow.

For complete details on the items listed above, refer to the [Infinium Assay Lab Setup & Best Practices Guide](https://support.illumina.com/downloads/infinium-assay-lab-setup-and-procedures-11322460.html) and the [Infinium Assay Consumables and Equipment List](https://support.illumina.com/downloads/infinium-consumables-equipment-1000000084294.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1507), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001507%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001507.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.