> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000008381.md).

# Troubleshooting Black Screen with EFI Shell message on the NovaSeq 6000

Users may find that upon starting the NovaSeq 6000, the initialization will stop and display a black screen with the following error:

EFI Shell version 2.31 \[4.654]

Current running mode 1.1.2

map: Cannot find required map name

Press ESC in 1 seconds to skip startup.nsh, any other key to continue.

Shell>

Shell>

Shell>

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-706dd10064651c188bdb011e18a2f5dcdfc01386%2Fimage1.jpg?alt=media\&token=9f675e95-617f-4668-970d-a4bed6c4f215)

This may indicate a communications issue onboard the instrument or may indicate problems with the onboard computer.

To resolve this error, perform a power cycle of the instrument using the power switch on the rear of the instrument. While this is typically not recommended, in this case the option for a normal power cycle will not be available.

1. Use the toggle switch at rear right to switch the instrument off.
2. Wait 5 minutes with the instrument completely shut down.
3. Use the toggle switch to turn the instrument on.
4. Once the power button on the right front side of the instrument begins to pulse, press the power button to turn the instrument computer on.
5. Log in as normal.

If there are any issues with re-initialization, or the subsequent run freezes or errors out, contact [Illumina Technical Support.](https://www.illumina.com/company/contact-us.html#/united-states/technical-support)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8381), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008381%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000008381.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.