# Troubleshooting Black Screen with EFI Shell message on the NovaSeq 6000

Users may find that upon starting the NovaSeq 6000, the initialization will stop and display a black screen with the following error:

EFI Shell version 2.31 \[4.654]

Current running mode 1.1.2

map: Cannot find required map name

Press ESC in 1 seconds to skip startup.nsh, any other key to continue.

Shell>

Shell>

Shell>

![](/files/Pmd92gMuLreIbN2nWhz4)

This may indicate a communications issue onboard the instrument or may indicate problems with the onboard computer.

To resolve this error, perform a power cycle of the instrument using the power switch on the rear of the instrument. While this is typically not recommended, in this case the option for a normal power cycle will not be available.

1. Use the toggle switch at rear right to switch the instrument off.
2. Wait 5 minutes with the instrument completely shut down.
3. Use the toggle switch to turn the instrument on.
4. Once the power button on the right front side of the instrument begins to pulse, press the power button to turn the instrument computer on.
5. Log in as normal.

If there are any issues with re-initialization, or the subsequent run freezes or errors out, contact [Illumina Technical Support.](https://www.illumina.com/company/contact-us.html#/united-states/technical-support)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8381), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008381%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000008381.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.