Troubleshooting "Component Failed" errors immediately after Pre Run Checks on the NovaSeq 6000
Background
The NovaSeq 6000 performs several key steps between the completion of the Pre-Run Checks and the start of a Sequencing run. If there are issues completing any of these steps, the Control Software typically stops the sequencing run within 1-2 minutes of completing the Pre-Run Checks and presents an error indicating, "A component failed during sequencing..."
If this issue is observed at the very start of the sequencing run, the consumables can often be re-used, though it will be important to review the issue with Illumina Tech Support before attempting a repeat sequencing run.
Troubleshooting Actions
If the Control Software presents, "A component failed during sequencing..." error within 1-2 minutes of starting the sequencing run, see the following steps and contact Illumina Tech Support.
Collect the following files/folders:
Control Software Logs: C:\ProgramFile\Illumina\NovaSeq\NovaSeq Control Software\Logs
Run Logs (if present): C:\ProgramData\Illumina\NovaSeq\NovaSeqTemp<RunID>\Logs
Universal Copy Service Logs: C:\ProgramData\Illumina\Universal Copy Service\Logs
Note whether the instrument is configured to connect to Proactive or BaseSpace
Select the dropdown menu in the top-left corner, then select Settings.
For BaseSpace, note whether the Cloud Settings or BaseSpace Sequencing Hub checkbox is selected.
For Proactive, note whether the Send Instrument Performance Data to Illumina checkbox is selected.
Once these files and information are collected, please call Illumina Tech Support for next steps.