Troubleshooting Cavro Pump Error on MiSeq

A Cavro Pump error can result from an issue with the instrument hardware or communication errors on the MiSeq. Potential hardware related causes include pump, pump controller, or syringe failure.

Error: Cavro pump error: Volume will exceed syringe capcity.Error in LanePump. Invalid argument. Command = '1V400IP1500R\r', reply = 'c'

Troubleshooting Steps:

  1. Power cycle\* the MiSeq. (Note: If a run is in progress, it will not be able to be resumed.)

  2. If there are no issues with re-initialization:

  • Perform post-run wash to ensure fluidics is operating normally.

  • Optionally, perform a volume test within the system check.

  1. If the instrument fails to initialize or there is an error completing the wash, contact Illumina Technical Supportarrow-up-right for further assistance.

For further information, search for the following Knowledge Base articles:* *\How to Shut down or power cycle the MiSeq.

  • *\How To Power Cycle the MiSeq Video.

  • System Checks for MiSeq.

  • Troubleshooting Lane Pump Errors on the MiSeq.

For any feedback or questions regarding this article (Illumina Knowledge Article #1454), contact Illumina Technical Support [email protected]envelope.