# Troubleshooting Cavro Pump Error on MiSeq

A Cavro Pump error can result from an issue with the instrument hardware or communication errors on the MiSeq. Potential hardware related causes include pump, pump controller, or syringe failure.

**Error**: Cavro pump error: Volume will exceed syringe capcity.Error in LanePump. Invalid argument. Command = '1V400IP1500R\r', reply = 'c'\
![](/files/TyKvoxfEfuHSUGjeQ2uA)

**Troubleshooting Steps**:

1. Power cycle\\\* the MiSeq. (Note: If a run is in progress, it will not be able to be resumed.)
2. If there are no issues with re-initialization:

* Perform post-run wash to ensure fluidics is operating normally.
* Optionally, perform a volume test within the system check.

3. If the instrument fails to initialize or there is an error completing the wash, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/) for further assistance.

For further information, search for the following Knowledge Base articles:\* \**\How to Shut down or power cycle the MiSeq.*

* \**\How To Power Cycle the MiSeq Video.*
* **System Checks for MiSeq.**
* **Troubleshooting Lane Pump Errors on the MiSeq.**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1454), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001454%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001454.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.