> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001521.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Simple guidelines for identifying top/bottom (TOP/BOT) strand and A/B allele

It can be challenging to determine the DNA strand and allele designations for a single nucleotide polymorphism (SNP) because strand designations and orientation can differ depending on the database or assembly referenced (eg, NCBI genome build updates).

To address this issue, Illumina developed the top/bottom (TOP/BOT) strand and A/B Allele nomenclature using sequence-based context to assign DNA strand orientation that allows the same allele calls over time regardless of database or genome build used.

**Unambiguous SNPs \[A/(G or C)] or \[T/(G or C)]**

For SNPs that are not \[A/T] or \[G/C], A is always on the top strand and T is always the bottom strand. A and T nucleotides are the Allele A”; G and C nucleotides are the Allele B”.

**Ambiguous SNPs \[A/T] or \[G/C]**

If the SNP is \[A/T] or \[G/C]: Use sequence walking to determine TOP/BOT strands, then assign A/B Alleles.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-57ab3fb1449323ad982382cbb566b03b32cfca94%2Fimage1.png?alt=media)

**Figure 1**. Sequence walking convention to determine TOP/BOT strands.

**How to use sequence walking to assign strands for ambiguous SNPs**

1. The ambiguous SNP position is “n” (Figure 1). Nucleotides one position upstream and one downstream from “n” are “n-1” and “n+1.” Nucleotides two positions upstream and two downstream from “n” are “n-2” and “n+2.” , etc.
2. Examine n-1|n+1. Is one of the pair either an “A” **or** “T” and the other a “G” **or** “C”?
   1. If no: Examine n-2|n+2. If needed, continue sequence walking until you find an n-*x*|n+*x* pairing in which one of the pair is either an “A” or “T” and the other is a “G” or “C.” Then proceed to Step b.
   2. If yes: Is the “A” or “T” in this unambiguous pair 5′ of the SNP position (“n”) or 3′ of the SNP position (“n”)?
      * If 5′: This is the TOP Strand.
      * If 3′: This is the BOT Strand.

**Note:** In Figure 1, an unambiguous pairing is present at n-1|n+1: \[C/T]. T is present 3' of the SNP positions, thus, this is the BOT Strand.

1. 1.

**How to assign nucleotide designations A or B Allele**

* **For \[A/T] SNPs**
  * when the Strand is TOP: Allele A = A and Allele B = T
  * when the Strand is BOT: Allele A = T and Allele B =A
* **For \[C/G] SNPS**
  * when the Strand is TOP: Allele A = C and Allele B = G.
  * when the Strand is BOT: Allele A = G and Allele B = C

**Note:** In Figure 1, the Strand is BOT and the SNP is \[A/T], thus Allele A = T and Allele B = A.

**Reference:**

[“TOP/BOT” Strand and “A/B” Allele](https://www.illumina.com/documents/products/technotes/technote_topbot.pdf)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1521), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001521%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001521.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.