Troubleshooting "Could not find a part of the path" errors on the NextSeq500/550 and MiniSeq
Users may encounter the following error during sequencing runs:
Could not find a part of the path 'D:\Illumina\ImagingMemoryDump\SparkMiniDump.dmp'.
An error occurred while sequencing. Will now proceed to determining whether to perform a post-run wash. Originally caused by: Could not find a part of the path 'D:\Illumina\ImagingMemoryDump\SparkM...."

This error results in the run stopping early and is non-recoverable.
Typically, this error is the result of a transient communication issue due to the Control Software being unable to access necessary files. Power cycling the instrument restores this connection and allows future runs to proceed without issue. There may also be an issue with the DMA Driver, so it is recommended to check the status of the DMA driver on the NextSeq 500/550 or MiniSeq as well.
Troubleshooting steps:
Check status of the DMA driver.
See instructions in the Illumina Knowledge Article Troubleshooting Bit Error on the NextSeq 500/550 and MiniSeq.
Power Cycle the instrument. Refer to the following instructions depending on the instrument:
If after power cycling the Control sSoftware fails to initialize or the subsequent run fails with the same error, contact Illumina Technical Support.
For any feedback or questions regarding this article (Illumina Knowledge Article #9036), contact Illumina Technical Support techsupport@illumina.com.