> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/on-premises-software/software-on-premises-software-reference_material-list/000009369.md).

# DRAGEN TruSight Oncology 500 (TSO500) co-install options

Co-installing DRAGEN TSO500 & DRAGEN TSO500 ctDNA

DRAGEN TSO500 v2.6.0 can be co-installed with one each of DRAGEN TSO500 and DRAGEN TSO500 ctDNA v2.5.x or v2.1.x as long as the packages use DRAGEN v3.10.x.

Scenarios for correct and incorrect co-installs:

DRAGEN TSO500 **v2.6.0** + DRAGEN TSO500 **v2.5.3**+DRAGEN TSO500 ctDNA **v2.5.0**: YES can be co-installed

DRAGEN TSO500 **v2.6.0** + DRAGEN TSO500 **v2.1.1** + DRAGEN TSO500 ctDNA **v2.5.0**: YES can be co-installed

DRAGEN TSO500 **v2.6.0** + DRAGEN TSO500 **v2.5.3** + DRAGEN TSO500 ctDNA **v2.1.1**: YES can be co-installed

DRAGEN TSO500 **v2.6.0** + DRAGEN TSO500 ctDNA **v2.5.0** + DRAGEN TSO500 ctDNA **v2.1.1**: NO cannot be co-installed

DRAGEN TSO500 **v2.6.0** + DRAGEN TSO500 DNA **v2.5.3** + DRAGEN TSO500 **v2.1.1**: NO cannot be co-installed

When co-installing, install DRAGEN TSO500 software versions in **ascending** order of the **DRAGEN software version** (as version below v2.6.0+ will overwrite with its corresponding DRAGEN Software version).

Example:

Co-install DRAGEN TSO500 **v2.1.1** (DRAGEN v3.10.9) with DRAGEN TSO500 **v2.6.0** (DRAGEN v3.10.17) steps:

1. Install DRAGEN TSO500 **v2.1.1**
2. Install DRAGEN TSO500 **v2.6.0**

If a prior version of DRAGEN TruSight Oncology 500 Analysis Software (eg. v2.5.3) is installed after v2.6.0, re-execute the installation script for v2.6.0 to install the compatible version of DRAGEN Software without impacting other installations.

**For more information about installation instruction, please check the** [**DRAGEN TSO500 v2.6 User Guide**](https://help.tso500software.illumina.com/dragen-tso-500-guides/dragen-tso-500-v2.6.0/getting-started/installation-on-standalone-dragen-server)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9369), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009369%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/on-premises-software/software-on-premises-software-reference_material-list/000009369.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.