Troubleshooting RheodyneValve error on NextSeq 500/550

The following error on the NextSeq 500/550 may indicate an issue with the sippers controlled by the valves. Rheodyne valve failure is uncommon, and this error is typically caused by a transient communication error.

Error in Fluidics.FlowCellFluidics1.LanePump.IxPassThroughRheodyneValve1: Reinit while recovering from error””. Originally caused by” Timeout waiting 60000ms within EventToBclockingConverter.WaitForData().

Troubleshooting steps:

  1. Power cycle the NextSeq 500/550, following the instructions in How to Power Cycle the NextSeq 500/550.

  2. From the System Checks, perform ONLY 'Valve Response Test'.

  3. Perform a manual quick wash for further testing.