Troubleshooting RheodyneValve error on NextSeq 500/550
The following error on the NextSeq 500/550 may indicate an issue with the sippers controlled by the valves. Rheodyne valve failure is uncommon, and this error is typically caused by a transient communication error.
Error in Fluidics.FlowCellFluidics1.LanePump.IxPassThroughRheodyneValve1: Reinit while recovering from error””. Originally caused by” Timeout waiting 60000ms within EventToBclockingConverter.WaitForData().

Troubleshooting steps:
Power cycle the NextSeq 500/550, following the instructions in How to Power Cycle the NextSeq 500/550.
From the System Checks, perform ONLY 'Valve Response Test'.
Select Manage Instrument > select System Check.
For further guidance on System Checks, follow How to perform a NextSeq 500/550 system check.
Perform a manual quick wash for further testing.
For further guidance, see How to perform washes on the NextSeq 500/550.
If the error does not recur, the instrument is clear for use.
If the error comes back during initialization, system check or during the wash, contact Illumina Technical Support.
For any feedback or questions regarding this article (Illumina Knowledge Article #1389), contact Illumina Technical Support techsupport@illumina.com.