Troubleshooting RheodyneValve error on NextSeq 500/550
Last updated
Was this helpful?
Last updated
Was this helpful?
The following error on the NextSeq 500/550 indicates that there is an issue regarding the sippers controlled by the valves. The Rheodyne valve is not prone to failure, and this typically results from a transient communication error.
Error in Fluidics.FlowCellFluidics1.LanePump.IxPassThroughRheodyneValve1: Reinit while recovering from error””. Originally caused by” Timeout waiting 60000ms within EventToBclockingConverter.WaitForData().
Troubleshooting steps:
Power cycle the NextSeq 500/550.
Run the ‘Valve Response Test’ portion of the System Check.
Select Manage Instrument > select System Check.
Perform a manual quick wash for further testing.
If the error does not recur, the instrument is cleared for use.
For further information, search for Knowledge Base articles How to Perform a NextSeq 500/550 Power Cycle and How to perform washes on the NextSeq 500/550.
If the error comes back during initialization, system check, or during the wash, contact .
For any feedback or questions regarding this article (Illumina Knowledge Article #1389), contact Illumina Technical Support .