> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000008466.md).

# How to troubleshoot Insufficient Memory errors during Pre-Run Checks on the NovaSeq X Series Instruments

**Background**

Before starting a run, the Pre-Run Checks verify that the instrument is ready to perform a sequencing run and are split between the Dry and Wet checks. If the instrument does not have sufficient memory available to perform all of the necessary software actions during these Checks, it will present the following error:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-236b4512ee550e22190cbd18758793f7a69a3eae%2Fimage1.png?alt=media\&token=c342c7ad-ee92-45e2-8539-29cd9c803174)

**Figure 1:** Pre-Run Checks Errors - Insufficient memory to continue the execution of the program.

If the Pre-Run Checks are in the Dry Checks phase when the error occurs, the consumables will not be marked as used and are safe to be used in a follow up sequencing run.

If the Pre-Run Checks are in Wet Checks phase when the error occurs, the Flow Cell and Buffer Cartridge will be marked as used and new Flow cell and Buffer consumables will be required for a follow up sequencing run.

If this issue is observed, see the following steps to troubleshoot this issue.

**Troubleshooting Actions**

**IMPORTANT NOTE**: this issue is distinct from insufficient storage space and deleting sequencing runs will NOT resolve this issue. For issues with insufficient storage, see [Troubleshooting Insufficient Storage Space Errors during Pre-Run Checks on the NovaSeq X Series](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007797.md).

1. Within the Control Software, select **Return to Start** to exit the Pre-Run Checks.
2. Power cycle the instrument according to Control Software version guidelines listed in the [How to Power Cycle the NovaSeq X Series Instruments](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007478.md) Knowledge Base Article.
3. Once the instrument has restarted and the Control Software is initialized, proceed with repeating the sequencing run setup.

If the issue persists, contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8466), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008466%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000008466.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.