Troubleshooting SIPDOWN Errors on the MiSeq

The MiSeq Control Software (MCS) will show a SIPDOWN error if the reagent sippers are not functioning or cannot lower to pierce the foil on the reagent cartridge.

  • This can be caused by physical obstruction (if the reagent cartridge is not fully inserted or is damaged), a hardware problem with the sipper assembly, or a communication error.

  • When present, this error occurs before the run starts, when the sippers first lower, immediately after selecting the start button.

  • This error can present in older versions of MCS as an Object Reference error.

Troubleshooting steps:

  1. Verify the reagent cartridge is fully inserted all the way to the back of the instrument chiller compartment.

  • Remove the reagent cartridge and verify no damage to the exterior of the cartridge. Confirm the reagents are fully thawed*.

    • Following a SIPDOWN error, the sippers are not lowered into the reagent cartridge. Open the door to remove the cartridge.

    • Alternatively, close and reopen the control software to exit the end of run sequence. Upon re-initialization, user may be prompted to perform post-run wash. Replace experimental flow cell and reagent cartridge with used flow cell and MiSeq wash cartridge for post-run wash.

  • Confirm if reagent foils are pierced.

  • Store the reagent cartridge, PR2 buffer, and the current flow cell in the buffer in the fridge (4°C).

  1. Power cycle** instrument to reset communications.

  2. After MCS re-initialization, select post-run wash.

  3. Perform the post-run wash with a previously used flow cell and the MiSeq wash cartridge.

  • Often, the run can be re-started with the same consumables. However, if those consumables are used to complete the post-run wash prior to reloading, the run will fail.

  • The post-run wash MUST be completed with previously used consumables and/or wash specific consumables.

  1. If there is no indication of damage to the reagent cartridge, and the post-run wash completes successfully, restart the run with the same consumables.

  2. If the reagent cartridge looks damaged, contact Illumina Technical Supportarrow-up-right and provide the reagent serial number (SN), cartridge lot number, and cartridge expiration date.

For further information, search for the following Knowledge Base articles:* Troubleshooting Sipdown Errors on the MiSeq Video.arrow-up-right

For any feedback or questions regarding this article (Illumina Knowledge Article #1456), contact Illumina Technical Support [email protected]envelope.

Last updated

Was this helpful?