# Cell input and volume loading recommendations for Illumina Single Cell 3’ RNA Prep

**Cell and Nuclei Loading Parameters**

Cell and nuclei loading in Illumina Single Cell 3â€™ RNA Prep kits is optimized to balance high recovery rates and data quality. The table below summarizes optimal loading parameters by kit size.

| Kit Size | Optimal Cell or Nuclei Input\* | Expected Capture Rate | Cell Loading Volume (µl) | RNase Inhibitor Volume (µl) | <p>Target<br>Concentration</p> |
| -------- | ------------------------------ | --------------------- | ------------------------ | --------------------------- | ------------------------------ |
| **T2**   | 5,000                          | ≥ 2,000               | 4                        | 1                           | 1,250 live cells/µl            |
| **T10**  | 17,000                         | ≥ 10,000              | 5                        | 1                           | 3,400 live cells/µl            |
| **T20**  | 40,000                         | ≥ 20,000              | 8                        | 1 - 2                       | 5,000 live cells/µl            |
| **T100** | 200,000                        | ≥ 100,000             | 20                       | 4                           | 10,000 live cells/µl           |

*\*Exceeding the recommended input increases the risk of a multiplet rate above 8%.*

**Notes:**

* Always include RNase inhibitor in the total loading volume.
* Refer to reference guides for loading volumes to ensure optimal capture efficiency.

**Cell and Nuclei Loading Volume** **Considerations**

* Adjust the target concentration if increasing or decreasing input.
* The total loading volume, including RNase inhibitor, must not exceed the specified limit for each kit size. Exceeding recommended volumes may reduce capture efficiency.
* If cells or nuclei cannot be concentrated to the target level, a lower concentration is generally acceptable as long as the total volume remains within recommended limits.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9473), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009473%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |