> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007797.md).

# Troubleshooting Insufficient Storage Space Errors during Pre-Run Checks on the NovaSeq X Series

**Background**

The NovaSeq X Series instrument is designed to output data to a networked storage location, though the data is stored on instrument temporarily until it is manually removed by the user. As part of the Pre-Run Checks, the instrument confirms that there is sufficient storage space on the instrument to complete both the Sequencing Run and the configured Secondary Analyses.

If there is insufficient space to complete these tasks, the Control Software presents a warning or error to the user and requires user input before proceeding. The specific messaging varies depending on Control Software version and the amount of space that is available.

**For Control Software v1.3 and above:**

If there is sufficient storage space to complete the Sequencing Run but not enough space for the configured Secondary Analyses, the software displays a Warning during the Pre-Run Check indicating that there is low disk space, but the sequencing run can still proceed.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8d2c72a945cdbe111f74f9b6b28dabbbabf92d52%2Fimage1.png?alt=media)

Pre-run checks warnings:

There is sufficient storage space to start sequencing, but not enough space to complete onboarding secondary analysis.

If you do not clear at least XXXX GB of storage space before sequencing completes, secondary analysis will be automatically canceled and you need to manually requeue the analysis after sufficient space is made available.

Options:

* Select **Back to consumables** to return to the consumable loading screen.
* Select **Return to Start** to return to the home screen.
* Select **Continue** to proceed with starting the run, however, if sufficient space has not been cleared before Sequencing completes, the Secondary Analysis is automatically canceled.
* Perform the actions under the Delete a Run section below to clear disk space during the Run to allow the Secondary Analyses to proceed. If the Secondary Analysis is cancelled, the user can use the Requeue workflow to restart the Secondary Analysis after the Sequencing Run completes.
  * See Knowledge Article [How to requeue or perform a new analysis of existing data on the NovaSeq X Series](https://github.com/illumina-swi/illumina-knowledge/tree/master/docs/Instrumentation/NovaSeq-X-X-Plus/Troubleshooting/000007765.md) for additional information.

**For the NovaSeq X Control Software v1.2.2 and below:**

If there is insufficient data storage capacity on the Compute Engine (CE) or Single Board Computer (SBC), regardless of whether it is for sequencing run or secondary analysis, the Control Software displays a Pre-Run Check Error that prevents starting the Sequencing Run.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-beccdd0842b55019cf21878c270b3aefec6834cd%2Fimage2.png?alt=media)

Pre-run checks errors:

There is insufficient storage space to start the run.

Navigate to the Runs screen to view current runs on the instrument, and delete enough runs on the \[SBC/CE] to clear at least XXXX GB of space.

Canceling one or both sides returns the respective side to the home screen. Use the following steps to check the available space consumed by Runs or Resources on the instrument, then clear space by deleting completed Runs from the instrument storage. Once sufficient space has been cleared, select **Retry** to proceed with the Pre-Run Checks or restart the run setup if it was canceled.

If the issue persists after clearing disk space by removing previous sequencing Runs or removing Resources, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

**Resolution Actions**

Check Disk Space

1. Select the **instrument icon** to open the global navigation menu.
2. Select **Runs**.
3. The available disk space and the disk space used by runs is in the upper right-hand area.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d198404d20af4fb9c0c862651ec4a82e7c5b70ae%2Fimage3.jpg?alt=media)

Disk Space Requirements

Estimates are for a 300 cycle, single-sided sequencing run and need to be doubled for a dual-sided run. Secondary analysis estimates are based on DRAGEN Germline App. Estimates vary depending on cycle number and DRAGEN Application.

| Flow Cell | SBC (GB) | CE Primary (TB) | CE Secondary (TB) |
| --------- | -------- | --------------- | ----------------- |
| 1.5B      | 15       | 0.4             | 1                 |
| 10B       | 19       | 2.1             | 5.5               |
| 25B       | 24       | 5.6             | 14.7              |

Note: The "CE Primary" refers to cBCLs file size generated, while the "CE Secondary" refers to the secondary analysis files generated by DRAGEN Apps.

Delete a Run

Only delete Runs using the NovaSeq X Series Control Software and do not delete Runs through the Linux Operating System (OS). Deleting runs manually through the Linux OS can negatively impact the control software. A deleted run cannot be requeued. Prior to deleting runs from the NovaSeq X Series Control Software, make sure the runs are backed up on a network drive or in BaseSpace Sequence Hub.

1. Select the **instrument icon** to open the global navigation menu.
2. Navigate to **Runs**.
3. For the run to delete, select the **ellipsis icon** in the **Action** column.\
   Choose one of the following options:
   * **Delete run data:** Deletes the sequencing and analysis output folders from the instrument, but does not remove the run from the Completed tab. The run details are still available in the control software, but the DRAGEN secondary analysis report is removed.
   * **Delete run:** Deletes the run data and removes the run from the Completed tab.
4. In the dialog box, confirm run deletion.
5. Repeat steps 3 - 5 to delete runs as needed.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d5531e988bd11fc0700cb58eece50989e719f135%2Fimage4.png?alt=media)

**Note:** These steps can also be performed using the Illumina Run Manager (EdgeOS) interface by opening the Chromium browser and navigating to <http://localhost>.

Remove Resources

Use the following instructions to remove reference genomes or reference files.

1. Select the **instrument icon** to open the global navigation menu.
2. Choose **Settings**, and then navigate to **Resources**.
3. Select the **Genomes** tab or **Reference files** tab.
4. For the genome or reference file to remove, select the **delete** icon.
5. In the dialog box, select **Yes, remove** to confirm.
6. Repeat steps 4 and 5 for each genome or reference file to remove. If needed, users can check the CE disk space prior to setting up a run.

If the issue persists after clearing disk space by removing previous sequencing runs, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7797), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007797%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007797.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.