> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000009854.md).

# Troubleshooting pre-run check failure due to "Imaging test failed" on the NextSeq 1000/2000

Users may encounter the following error during pre run checks:

**Imaging test failed. The imaging metric Worst bit error rate was out of spec.**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f2304d93f43236d3d264f13ea3e9852bf8abc362%2Fimage1.jpg?alt=media)

This error occurs during dry pre run checks, reagents have not yet been marked as used.

**To troubleshoot**:

1. Select "Retry" to re-attempt starting the run.
2. If error persists after retrying, Power Cycle the instrument and attempt to start the run again. See [How to power cycle the NextSeq 1000/2000](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003038#:~:text=To%20power%20cycle%20the%20instrument%3A%20From%20the%20control,side%20of%20the%20instrument%20until%20the%20lights%20fade.).
3. If error persists after power cycling instrument, collect the following files and contact Illumina Technical Support:
   1. Perform **Bit Error Rate test** of System Check, results are found within `/usr/local/illumina/systemcheck`. See [System Checks for the NextSeq 1000/2000](https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-reference_material-list/000003457).
   2. **Control Software logs** within `/usr/local/illumina/logs/next1k2k-control-software`
   3. **Run** **set up logs** within `/usr/local/illumina/runs/setup/`

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9854), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009854%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-1000-2000/instrumentation-nextseq-1000-2000-troubleshooting-list/000009854.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.