> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009774.md).

# Troubleshooting '\[XYZ] encountered an exception on the IX server' errors on the MiSeq i100 Series

**Background**

The MiSeq i100 Series establishes an internal network to connect the onboard computer with the control boards on the instrument. If the connection between the computer and a board is interrupted, the instrument Control Software displays the following Critical System Error:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6f7099500cfa2903ed5973b29cce9ca474eac7aa%2Fimage1.png?alt=media)

`Critical System Error: '[XYZ] encountered an exception on the IX server'`

In this error message, `[XYZ]` is the acronym for the specific control board that lost connection to the onboard computer.

If this error presents, please see the following **Troubleshooting Actions**.

**Troubleshooting Actions**

1. On the Critical System Error page, select **Shutdown** to power down the instrument.
   1. Once powered down, toggle the **power switch** on the back of the instrument to **OFF** and leave instrument off for 2 minutes.
   2. Toggle the power switch to **ON**, then press the power button on the **front-right corner** of the instrument.
   3. See Knowledge Article [How to Power Cycle the MiSeq i100 Series](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009277.md) for additional details if needed.
2. After instrument boots, monitor **Control Software** for successful initialization.
3. If instrument initializes as expected, continue with a normal workflow.

If the instrument fails to initialize and continues to present `Critical System Error: '[XYZ] encountered an exception on the IX server'` messages, contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9774), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009774%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009774.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.