Following the NextSeq 500/550 Windows 10 upgrade, Hard Drive swap or LCD Screen replacement, the touch screen may respond to inputs an unexpected manner. Touch inputs may be inverted. For example, selection at the bottom right hand corner of screen results in click at top left of screen or inaccurate screen calibration.
To resolve this:
From the Windows Start menu, search for "PenMount Control Panel"
Open the PenMount Control Panel
Select PenMount 6000 USB from the main window
Select Configure
In the new window, open the Calibrate tab, then select Standard Calibration
Follow directions as prompted
When the control window reappears, select OK, then OK again in the next screen to exit
Confirm the touch screen is now responding as expected
If issue persists, contact Illumina Technical Support for assistance
For any feedback or questions regarding this article (Illumina Knowledge Article #7990), contact Illumina Technical Support techsupport@illumina.com.