Troubleshooting "Instrument Shell has Stopped Working" on the NovaSeq 6000
Last updated
Was this helpful?
Last updated
Was this helpful?
The "Instrument Shell has Stopped Working" error on the NovaSeq 6000 is a Windows error that generally indicates transient communications issues or hardware problems. This error will manifest as the Control Software crashing, and the pop-up shown below appearing on the screen.
Instrument Shell has stopped working. A problem caused the program to stop working correctly. Please close the program.
Close the program
Debug the program
To Troubleshoot:
Power cycle the instrument. For further information, search for Knowledge Base article How to Power cycle the NovaSeq 6000 System.
Collect the following files:
NovaSeq Control Software logs folder: C:\ProgramData\*\Illumina\NovaSeq\NovaSeq Control Software\Logs (zipped)
Windows logs - After collecting the logs, contact Illumina Technical Support (techsupport@illumina.com) to determine the next steps.