Troubleshooting Pressure decay or Impedance test stopped with NaN results on the NovaSeq 6000
Last updated
Was this helpful?
Last updated
Was this helpful?
If Pressure Decay or Impedance test on the NovaSeq 6000 system check stops with NaN result:
Check pressure sensor readings:
Close NovaSeq Control Software.
Open NovaSeq Service Software (NVSS).
Navigate to Integrated Chassis Assembly.
Navigate to the Sensors Tab and select the refresh icon for Fluidics.
If any sensor is >2.5psi, take a screenshot or picture of the result and collect NovaSeq Service Software Logs located under C:\ProgramFile\Illumina\NovaSeq\NovaSeq Service Software\Logs and contact Illumina Technical Support for the next steps.
Note: Pressure sensors are not used during a run and, therefore, failing sensors will not impact a run performance. The instrument can continue sequencing until an Illumina Field Service Engineer (FSE) can replace the sensors.