Troubleshooting Wash State Unknown in NextSeq 500/550 Control Software (NCS) v4.0.0
Last updated
Was this helpful?
Last updated
Was this helpful?
The following warning can appear in NextSeq Control Software (NCS) v4.0.0 after the instrument is rebooted and initialized. The NextSeq Control Software prompts a notification mentioning the wash state is unknown and then allows for either proceeding with a wash or no wash. This error is specific to NCS 4.0.0 and was fixed in Control Software update v4.0.1.
Troubleshooting steps:
Update the NextSeq Control Software from v4.0.0 to v4.0.1.
Contact Illumina Technical Support for any additional questions.