Universal Copy Service failed to validate the Output Folder error on the NextSeq 1000/2000
Last updated
Was this helpful?
Last updated
Was this helpful?
Universal Copy Service (UCS) copies sequencing output files from the run folder to BaseSpace Sequence Hub (if applicable) and the output folder.
Error Language: Universal Copy Service failed to validate the Output Folder. Retry and, if error persists, contact Illumina Technical Support.
Figure 1: Example of error in NextSeq 1000/2000 control software.
Troubleshooting steps:
The NextSeq 1000/2000 Control Software requires at least two levels of nested folders to recognize the location of the output folder on the mounted network drive/external portable drive.
Make sure that the information in the fstab file is correct. (ie, group-level permissions are set correctly, and no syntax errors are present).
For any feedback or questions regarding this article (Illumina Knowledge Article #3547), contact Illumina Technical Support techsupport@illumina.com.