Uninterruptible Power Supply (UPS) Models by Instrument
UPS information for Illumina Systems
Instrument
Included?
Brand (North America)
Brand (Japan)
Brand (International)
Site Prep Link (Page)
iSeq 100
No
APC Smart-UPS 750 VA LCD 120 V US
APC Smart-UPS 750 LCD 100 V
APC Smart-UPS 750 VA LCD 230 V
MiniSeq
No
APC Smart UPS 2200 VA LCD 120 V
APC Smart UPS 2200 VA LCD 120 V
APC Smart UPS 2200 VA LCD 230 V
MiSeq
No
APC Back-UPS Pro
#BR1500MS2
APC Smart UPS #SUA1500JB
APC Back-UPS Pro
#BR1500MS2
NextSeq 500/550
No
APC Smart UPS 1500 VA LCD 120 V
APC Smart UPS 1500 VA LCD 100 V
APC Smart UPS 1500 VA LCD 230 V
NextSeq 1000/2000
No
APC Smart UPS 1500 VA LCD 120 V
APC Smart UPS 1500 VA LCD 100 V
APC Smart UPS 1500 VA LCD 230 V
HiSeq 2500 series
Yes
APC SmartUPS 2200VA Model SUA2200JB
APC SmartUPS 2200VA Model SUA2200XLI
HiSeq 3000/4000
Yes
APC SmartUPS 2200VA Model SUA2200JB
APC SmartUPS 2200VA Model SUA2200XLI
HiSeq X
Yes
APC SmartUPS 2200VA Model SUA2200JB
APC SmartUPS 2200VA Model SUA2200XLI
NovaSeq 6000
Yes
Worldwide model - some countries excluded, see site prep guide
Worldwide model - some countries excluded, see site prep guide
NovaSeq X/X Plus
Yes
Worldwide model - some countries excluded, see site prep guide
Worldwide model - some countries excluded, see site prep guide
iScan
Yes
ABCE802-11B
ABCE800-11B
ABCE800-22B
Tecan
Yes
APC-SMT1500C
ABCE800-11B
ABCE800-22B
Important notes:
Illumina recommends that instruments be connected to the UPS, and that the UPS be connected to emergency power in the lab/institution.
UPSs are not meant to support runs for long periods of time - they are only meant to get the instrument over the initial outage before emergency power is connected (so only for a manner of minutes at most).
Use one UPS per instrument (do not plug in multiple instruments into one UPS) - multiple instruments/computers on a single UPS can cause spontaneous power failures.