> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/local-run-manager/software-local-run-manager-reference_material-list/000001317.md).

# How to use a custom library prep kit in Local Run Manager v2

[Local Run Manager](http://support.illumina.com/sequencing/sequencing_software/local-run-manager.html) is the integrated run planning and analysis software for Illumina benchtop sequencers. Version 2 is available for off-instrument analysis of run data from iSeq 100, MiniSeq, MiSeq, and NextSeq 500/550 systems. It is also available on-instrument for the iSeq 100 and MiSeq instruments running MiSeq Control Software version 3. Runs can be configured using sample sheets generated by Illumina software such as [Illumina Experiment Manager](https://support.illumina.com/sequencing/sequencing_software/experiment_manager.html) or through the Local Run Manager interface. Standard Illumina library prep kits are included as built-in options in Local Run Manager. Version 2 also includes a Custom option for manual entry of sample and index information. Custom and third-party library prep kits can also be added, and must be uploaded to Local Run Manager via a custom library prep kit file prior to setting up a run. The [Local Run Manager Software Guide](https://support.illumina.com/sequencing/sequencing_software/local-run-manager/documentation.html?langsel=/us/) contains information on the content of the necessary custom library prep kit file, and this article provides more details.

**Important note:** These directions are intended for version 2 of Local Run Manager only.

**Important note:** Although iSeq 100 utilizes Local Run Manager 2 and is capable of sequencing using custom primers with the release of the [iSeq Control Software version 3.1](https://support.illumina.com/downloads/iseq-system-software-v31.html), runs cannot be set up in Local Run Manager when using custom primers. iSeq 100 runs with custom primers must be set up in Manual mode on the instrument as described in the [iSeq 100 Sequencing System Documentation](https://support-docs.illumina.com/IN/iSeq100/Content/IN/FrontPages/iSeq_100_Sequencing_System.htm).

To add a custom library prep kit to Local Run Manager 2, perform the following steps:

1. Download the custom kit template from the Local Run Manager [Downloads page](http://support.illumina.com/sequencing/sequencing_software/local-run-manager/downloads.html) or by opening the Library Prep Kits page from the Local Run Manager **Tools** menu and selecting **Download Template**. The templates for existing kits can also be downloaded by selecting the download button next to that kit name.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a5f435891a2e252eae2ed6433ada1cf936c5ab67%2Fimage1.jpg?alt=media\&token=a3bdb945-88c5-4b57-aa9b-a6fed17c888d)
2. Open the template file in either Excel or a plain text editor, such as Notepad and edit the fields as follows:
   1. **\[Kit]**: This section contains the default run parameters:
      1. The **Name** section must be unique for all kits.
      2. **Description**: A short description of the kit. Only alphanumeric, period, dash, and space characters are allowed.
      3. Available options for **IndexStrategy** are NoIndex, SingleOnly, DualOnly, SingleAndDual, NoAndSingle, NoAndDual, or All.
      4. Available options for **ReadType** are Single, Paired, or All.
      5. The default number of cycles for Read 1 and Read 2 are specified in **DefaultReadLength1** and **DefaultReadLength2**, respectively.
   2. **\[Resources]**: Set custom parameters here. This section is optional. For more information on available parameters, see the "Library Prep Kits" section of the [Local Run Manager Software Guide](https://support.illumina.com/sequencing/sequencing_software/local-run-manager/documentation.html?langsel=/us/).
      1. **Adapter trimming (optional):** [Adapter sequences](/library-preparation/general/library-preparation-general-reference_material-list/000001314.md) can be included in the **Resources** section of the kit definition to enable adapter trimming of FASTQs; adapter trimming can also be toggled off during run setup in the **Module-Specific Settings**. For paired end kits with the same adapter on both reads (such as Nextera-based kits), only a single Adapter entry is needed. For kits where the Read 1 and Read 2 adapter sequences differ (such as TruSeq-based kits), a separate AdapterRead2 entry is needed. An example of how to enter these values is shown for a TruSeq kit below, where the arrows indicate tab entries in the TSV file:\
         ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c3b5be9a3c1d356acc0bfd3b7b70fcc7bcebb868%2Fimage2.jpg?alt=media\&token=2289c0e1-417e-4c28-a36f-d146c9b54b18)
   3. **\[Indices]**: Specify which index sequences will be used with this kit. Include the **Name** along with the **Sequence** and **IndexReadNumber** (1 or 2). After loading the custom kit into Local Run Manager, sequences are selected using the **Name** value.\
      **Important Notes**:
      1. For MiniSeq and NextSeq 500/550 users, do not reverse complement the second index sequence. Local Run Manager will do so automatically during demultiplexing as necessary based on the instrument type.
      2. Mixing indexing lengths and types (eg, dual and single) within a kit and run is not supported. Index lengths must be consistent and only A, G, C, and T bases are valid. N cannot be entered and used as a wildcard in a library prep kit, but can be manually entered when using the Custom option.
      3. Index names are limited to 50 characters. Attempting to upload a kit template with index names longer than 50 characters will result in an error on import.
   4. **\[SupportedModules]**: Include the name of the analysis module or modules which can make use of this kit when creating a run. Multiple modules should be included on separate lines. The full module name must be used, with no spaces, such as GenerateFastQWorkflow. Refer to the [Local Run Manager Software Guide](https://support.illumina.com/sequencing/sequencing_software/local-run-manager/documentation.html?langsel=/us/) for a full list of available modules.
3. Save the file, making sure to retain the format as TSV. Do not include spaces or special characters in the file name.
4. From the Local Run Manager dashboard, select the **Tools** dropdown menu, **Library Prep Kits**, then **Add Library Prep Kit**. Browse to the location of the custom kit file to add it.
5. From the dashboard, create a new run. In the Run Settings section, select the custom kit from the Library Prep Kit dropdown menu and then enter the read length information.
6. Continue with entering the sample information and complete setting up the run.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1317), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001317%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/software/local-run-manager/software-local-run-manager-reference_material-list/000001317.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
