> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000003817.md).

# Troubleshooting demultiplexing issues for local DRAGEN analysis onboard the NextSeq 1000/2000

DRAGEN software outputs two demultiplexing files called Demultiplex\_Stats.csv and Top\_Unknown\_Barcodes.csv that can be used to troubleshoot demultiplexing issues.

**Where are the Demultiplex\_Stats.csv and Top\_Unknown\_Barcodes.csv files located?**

\[runfolder]/Analysis/1/Reports

**What information is in the Demultiplex\_Stats.csv?**

The Demultiplex\_Stats.csv file provides the following information:

* Sample ID
* Index: index combination
* ## Reads: number of reads
* ## Perfect Index Reads: number of perfect index reads
* ## One Mismatch Index Reads: number of one mismatch index reads
* ## of >= Q30 Bases (PF): number of bases greater than or equal to Q30 (PF)
* Mean Quality Score (PF)

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-78915941808e142f20188ea22a976522617d53ec%2Fimage1.png?alt=media\&token=d94edfe9-a8f0-4a46-b395-bc9833122337)

**What information is in Top\_Unknown\_Barcodes.csv?**

The index and index2 columns provide the sequence of the unknown barcode as it was read by the instrument. The # Reads column provides the number of reads that the combination of index and index2 received.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-0619a9d7419c1626713ff671c624cb238801dbc4%2Fimage2.png?alt=media\&token=f11c3106-2dfc-4ef5-a80c-360ce1e66993)

When troubleshooting demultiplexing issues, use the index sequences identified for comparison to the expected index sequences (found in the sample sheet).

Possible causes for poor demultiplexing:

* Index sequences entered in the incorrect orientation in the sample sheet.
* Incorrect index sequences entered in the sample sheet (eg, Nextera vs TruSeq HT, or index A001 vs index A006).
* Poor Index Read sequencing quality (e.g., index sequences containing lots of periods, representing “no calls,” near the top of the list).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3817), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003817%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000003817.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.