> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000003577.md).

# How to demultiplex an incomplete run using bcl2fastq2 and BCL Convert

**How to demultiplex an incomplete run using bcl2fastq2**

Add the use-bases-mask and ignore-missing flags to the bcl2fastq2 command to generate FastQ files from an incomplete run.

`--use-bases-mask`

Arguments:

* N or n - Ignore the cycle.
* Y or y - Use the cycle.
* I or i - Use the cycle for an Index Read

For example, if a run set up is 2x100 cycles with dual 10 bp index reads and the run aborted at cycle 50 of the second read, add the following flag to demultiplex the run:

`--use-bases-mask Y*,I*,I*,Y49N*`

In this example, all the cycles of Read 1, all the index cycles, and 49 cycles of Read 2 are used while the rest of the cycles are masked out.

**Note**: use one fewer cycle than completed by the run.

In addition, add the following flags as well:

* `--ignore-missing-bcls`: The software ignores missing or corrupt BCL files and assumes 'N'/'#' for missing calls.
* `--ignore-missing-filter`: The software ignores missing or corrupt filter files and assumes that all clusters in tiles with missing filter files passed filter.
* `--ignore-missing-positions`: The software ignores missing or corrupt cluster location files. When cluster location files are missing, the software writes unique coordinate positions into the FASTQ header.

**How to demultiplex an incomplete run using BCL Convert**

* Determine the last cycle sequenced.
* Create a new v2 sample sheet and use the OverrideCycles setting to mask the un-sequenced cycles using "N".
  1. For example, if a run was set up as 2x151 cycles with dual 8 bp index reads and the last cycle sequenced was 177, set OverrideCycles to Y151;I8;I8;Y10N141.
  2. For more information, see [How to use the 'OverrideCycles' setting in the Run Planning User Interface](https://knowledge.illumina.com/software/general/software-general-faq-list/000005671.md).
* If the run aborted during the index cycles, use the OverrideCycles setting to mask the un-sequenced cycles using "N" and edit the sample sheet to remove the bases of the index that were not sequenced.
  1. For example, if a run was set up as 2x151 cycles with dual 8 bp index reads and the last cycle sequenced was 162, set OverrideCycles to Y151;I8;I3N5;N151 and remove the last 5 bases from the index2 column in the sample sheet.
     1. Note: Reducing the number of index bases will impact the ability to demultiplex and may result in barcode collision. Allowing zero barcode mismatches may be required to allow analysis to proceed.
* Upload the new sample sheet to a Project if using the [BCL Convert app in BaseSpace](https://www.illumina.com/products/by-type/informatics-products/basespace-sequence-hub/apps/bcl-convert.html).
* Requeue analysis using BCL Convert.

**Note**: BCL Convert, by default, has its strict mode turned off (`--strict-mode false`), so missing BCL/filter/BCI/locs files do not result in the analysis erroring nor ending prematurely.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3577), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003577%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000003577.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.