The information in this article applies to run data that was demultiplexed on BaseSpace Sequence Hub with the FASTQ Generation application, and to run data that was demultiplexed using the bcl2fastq2 conversion software version 2.17 and later. The information below is most applicable for the following instrument types that are set up using v1 Sample Sheets: MiSeq, NextSeq 550, iSeq, and MiniSeq.

After demultiplexing, the BaseSpace Sequence Hub FASTQ Generation application and the bcl2fastq2 v2.17+ conversion software will output a demultiplexing summary file called DemuxSummaryF1L#.txt. The L# stands for the lane number on the flow cell, and one summary file is output for each lane. These DemuxSummaryF1L#.txt files can be used to troubleshoot demultiplexing issues.

Where are the DemuxSummaryF1L#.txt files?

  • bcl2fastq2 v2.17+ After FASTQ file generation completes, the DemuxSummaryF1L#.txt files are located in the Stats folder, which is located in the specified output directory.

BaseSpace Sequence Hub For runs submitted to BaseSpace with a v1 sample sheet, the DemuxSummaryF1L#.txt files are located in the project after FASTQ file generation completes. To find the files in your BaseSpace Hub project:

    1. Navigate to the run page, and select the FASTQ Generation link under the "Latest Analyses" (Figure 1).

Figure 1: The Run Summary Page that lists the FASTQ Generation Analysis

    1. On the Analysis Info page, scroll down to the Log Files (Figure 2). You will find the DemuxSummaryF1L#.txt files on the list of log files.

  • Figure 2: The log files list in the Analysis page.

What information is in the DemuxSummaryF1L#.txt files?

  • The DemuxSummaryF1L#.txt file has 2 sections. The top section (Figure 3) is a tab-delimited table that summarizes the demultiplexing results on a per-tile basis. On the left side of the table is a list of the tiles on the flow cell lane. Along the top of the table, the samples are listed in the order in which they were entered into the sample sheet. Sample 0 is always reserved for the undetermined reads. The table shows the percentage of reads demultiplexed to each sample, per tile. In general, for a given sample, the percentage of demultiplexed reads should be similar across all tiles. The tile summary information can be used to identify tile-specific demultiplexing issues.

Figure 3: The tile summary section shows the percentage of reads demultiplexed to each sample, per tile.

The second section of the DemuxSummaryF1L#.txt file (Figure 4) lists the 1,000 most common undetermined index sequences, and the number of times each index sequence was found (or the number of clusters assigned to each index sequence).

Figure 4: The most popular index sequences section of the DemuxSummaryF1L#.txt file lists the top 1000 index sequences associated with the undetermined reads.

When troubleshooting demultiplexing issues, this list can be used to compare the expected index sequences (those in the sample sheet) to those that were found. Some common causes for poor demultiplexing that these lists can reveal are:

  • Index sequences entered in the wrong orientation in the sample sheet.

  • Incorrect index sequences entered in the sample sheet (eg, Nextera vs TruSeq UD or index A001 vs index A006).

  • Sample mix ups between lanes.

  • Poor Index Read sequencing quality.

    • Ns in the sequences represent positions where the base calling software was unable to make a base call.

    • For sequencing systems using Illumina one-channel (iSeq) or two-channel SBS chemistry (MiniSeq, NextSeq 500/550 and NovaSeq), poly-G sequences indicate that no index sequence was read. Poly-G sequences are typical for PhiX reads, which are not indexed.

For any feedback or questions regarding this article (Illumina Knowledge Article #2875), contact Illumina Technical Support techsupport@illumina.com.