> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000002415.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# How to remove a BCL file from analysis in MiSeq Reporter or Local Run Manager

**Background**

In situations where a single BCL file is corrupted or displays a high error rate, it is possible to remove the affected BCL from secondary analysis. This removes the affected tile at the impacted cycle from all downstream analysis.

Instructions:

1. Using Sequencing Analysis Viewer (SAV), identify the affected tile(s) by mouseover the affected part in the Flow Cell Chart.
2. On MiSeq, open `D:\Illumina\MiSeqAnalysis\<run folder>\Data\Intensities\BaseCalls\L001`
3. Open the folder for the affected cycle of the run.

a. For example, for cycle 123, the folder name is C123.1

4. Locate the BCL file for the affected flow cell tile.

a. For example, for tile 2103, the BCL file name is s\_1\_2103.bcl

5. Append .bak to the BCL file name.

a. For example, s\_1\_2103.bcl.bak

6. Requeue analysis of the run to repeat analysis with the problematic BCL file excluded from the results. For all of the clusters on the affected tile in the affected cycle, the base calls will be replaced with an N (no call).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2415), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002415%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000002415.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.