> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000002227.md).

# Troubleshooting demultiplexing issues using MiSeq Reporter

The [MiSeq Reporter](http://support.illumina.com/sequencing/sequencing_software/miseq_reporter.html) analysis software outputs a demultiplexing summary file called DemultiplexSummaryF1L1.txt that can be used to troubleshoot demultiplexing issues.

**Where is the DemultiplexSummaryF1L1.txt file?**

* After FASTQ file generation completes, the DemultiplexSummaryF1L1.txt file is located in:\
  \<run folder>\Data\Intensities\BaseCalls\Alignment.

**What information is in the DemultiplexSummaryF1L1.txt file?**

* The DemultiplexSummaryF1L1.txt file has 2 sections. The top section (Figure 1) is a tab-delimited table that summarizes the demultiplexing results on a per-tile basis. On the left-hand side of the table is a list of the tiles on the flow cell. Along the top of the table, the samples are listed in the order in which they were entered into the sample sheet. Sample 0 is always reserved for the undetermined reads. The table shows the percentage of reads demultiplexed to each sample, per tile. The tile summary information can be used to identify tile-specific demultiplexing issues.

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-162499f142464ec6b9b06eb22f922eee3d82acd5%2Fimage1.png?alt=media\&token=350ec36c-66ea-4c8f-85c5-d92798967716)

**Figure 1**. The tile summary section shows the percentage of reads demultiplexed to each sample, per tile.

The second section of the DemultiplexSummaryF1L1.txt file (Figure 2) lists the 100 most common index sequences (whether assigned or undetermined) for Index Read 1 and Index Read 2, and the number of times each index was found. The list also shows the reverse complement of the index sequences found (for reference only). In the DemutiplexSummaryF1L1.txt file, periods (...) in the index sequences represent positions where the base calling software on the MiSeq could not make a base call.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-b66b54a44a59d321753a2f70802c1187b43b19b4%2Fimage2.png?alt=media\&token=3cacfa9b-e837-4d19-af15-f5258d53df47)

**Figure 2**. The most popular index sequences section of the DemultiplexSummaryF1L1.txt file lists the top 100 index 1 and index 2 sequences found.

When troubleshooting demultiplexing issues, the lists of index sequences can be used to compare the expected index sequences (those in the sample sheet) to those that were found.

Some common causes for poor demultiplexing that this list can reveal are:

* Index sequences entered in the incorrect orientation in the sample sheet.
* Incorrect index sequences entered in the sample sheet (eg, Nextera vs TruSeq, or index A001 vs index A006).
* Poor Index Read sequencing quality (eg, index sequences containing lots of periods, representing "no calls," near the top of the list).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2227), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002227%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000002227.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.