> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general/software-general-reference_material-list/000001325.md).

# How to generate FASTQ files from an incomplete run in MiSeq Reporter

It is possible to generate FASTQ files, even if a MiSeq run does not complete. To do so, determine the last cycle with a complete set of \*.bcl files, modify run files, and requeue analysis. This bulletin describes how to generate FASTQ files from an incomplete run using MiSeq Reporter. If you are using Local Run Manager instead of MiSeq Reporter see the bulletin Local Run Manager: [Generate FASTQs from an incomplete run in Local Run Manager](https://knowledge.illumina.com/software/general/software-general-troubleshooting-list/000001807.md).

**Before starting**

* If you plan to do secondary analysis on the MiSeq, make sure that a sequencing run is not in progress and that MiSeq Reporter is ready for analysis.
* If you are using BaseSpace Sequence Hub for analysis and FASTQ generation instead of MiSeq Reporter, email <techsupport@illumina.com> for assistance.
* If the run stopped before completing the index reads (ie, in Read 1), it is not possible to demultiplex your samples.
* Make sure that the **RTAComplete.txt** file is **not** present in the run folder.

**Steps to take:**

1. Navigate to **D:\Illumina\MiSeqAnalysis\[runfolder]\Data\Intensities\BaseCalls\L001** and check which cycle was the last to generate all \*.bcl files. If v3 reagents were used, there are 38 \*.bcl files per cycle. If v2 reagents were used, there are 28 \*.bcl files per cycle.

\*If .bcl files are not present, it might be possible to recover them by manually launching RTA (Real-Time Analysis). To do so, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

Calculate how many cycles have been successfully completed and extracted.

* For example:
  * Run stopped at cycle 207 in a 2 x 150 + 6 bp index run. If the last cycle with complete \*.bcl files was cycle 206, it is a good rule of thumb not to push against the last cycle with complete \*.bcl files. Select the previous complete cycle from which to do the analysis. In this case, it is cycle 205 . The run breakdown is (R1) 150 + (R2 Index) 6 + (R3) 49 = 205 cycles.

2. Backup the original **RunInfo.xml** and **SampleSheet.csv** files by renaming them to something like **RunInfo.xml.BAK** and **SampleSheet.csv.BAK**. Using Notepad, edit these documents and save the edited documents as **RunInfo.xml** and **SampleSheet.csv** in the root run folder in MiSeqAnalysis. If the reverse read of a paired-end run - Read 3 in the following example - has not been performed, delete the corresponding line in **RunInfo.xml** and **SampleSheet.csv**.

* For Example:

RunInfo.xml:\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d4de1a3eadd6ae0b64780636e0df523fd71bc979%2Fimage1.jpg?alt=media\&token=5ad4857f-c7ee-4647-9efa-15691190db87)

SampleSheet.csv:\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-fdb601d964f9e6b2e0a8c70cd9d6b98fffe57466%2Fimage2.png?alt=media\&token=b0bdf3c6-a1a0-4360-a730-d201167cb6ae)

3. Copy an **RTAComplete.txt** file from a successful run folder and paste it into your run folder.
4. You can expect to see the run queued in MiSeq Reporter and a **QueuedForAnalysis.txt** file created in the run folder in MiSeqAnalysis. To make sure that the run has been requeued in MiSeq Reporter, point a web browser window to <http://localhost:8042> and look for your run in the list opened with the **Analyses** button.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1325), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001325%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general/software-general-reference_material-list/000001325.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
