> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general-library-prep/software-general-library-prep-reference_material-list/000002905.md).

# Adapter trimming - Why are adapter sequences trimmed from only the 3' ends of reads

Illumina FASTQ file generation pipelines include an adapter trimming option for the removal of adapter sequences from the 3' ends of reads. Adapter sequences should be removed from reads because they interfere with downstream analyses, such as alignment of reads to a reference. The adapters contain the sequencing primer binding sites, the index sequences, and the sites that allow library fragments to attach to the flow cell lawn. Libraries prepared with Illumina library prep kits require adapter trimming only on the 3' ends of reads, because adapter sequences are not found on the 5' ends.

To understand why adapter sequences are found only on the 3' ends of the reads, it helps first to understand where the sequencing primers anneal to the library template on a flow cell. The diagrams below show the sites of primer annealing at each stage of sequencing run: Read 1, Index 1, Index 2 and Read 2.

Figure 1. MiSeq and NovaSeq 6000 (with v1 reagents)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d9099352b312adb3c3f15560cfab0744d9564ffa%2Fimage1.png?alt=media\&token=2d294364-12d7-427e-9981-9948442f6b63)

Figure 2. iSeq 100, MiniSeq, NextSeq 500/550, NextSeq 1000/2000, NovaSeq 6000 (with v1.5 reagents), NovaSeq X Series, and MiSeq i100 Series (with read first run configuration)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ea0ae9274953a3714c4a1df69dbba25b2484deff%2Fimage2.png?alt=media\&token=92f5723e-68cb-4eb6-b579-d635096b3699)

As shown in Figures 1 and 2, in both Read 1 and Read 2, the sequencing primer anneals to the adapter, immediately upstream of the DNA insert (in gray). Because the sequencing starts at the first base of the DNA insert in Reads 1 and 2, the adapter is not sequenced at the start of the read. However, if the sequencing extends beyond the length of the DNA insert, and into the adapter on the opposite end of the library fragment, that adapter sequence will be found on the 3' end of the read. Therefore, reads require adapter trimming only on their 3' ends.

**Additional information**

* The recommended sequences to use for adapter trimming for Illumina library prep kits are found in [Sequences to use for adapter trimming](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001314.md).
* The full adapter sequences for Illumina library prep kits are found in the [Illumina Adapter Sequences Document](https://support.illumina.com/downloads/illumina-adapter-sequences-document-1000000002694.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2905), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002905%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general-library-prep/software-general-library-prep-reference_material-list/000002905.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.