> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/bclconvert-bcl2fastq/software-bclconvert-bcl2fastq-reference_material-list/000008935.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# How to set OverrideCycles values for variable index sequence lengths on the NovaSeq X Series

BCL Convert 4.1+ allows a per-sample OverrideCycles value for each sample in the sample sheet \[BCLConvert\_Data] section. Below is an example of some [per sample settings](https://support-docs.illumina.com/SW/dragen_v42/Content/SW/DRAGEN/SampleSheet.htm?Highlight=%22per%20sample%22) that can be used in this section:

`[BCLConvert_Data]Lane,Sample_ID,Index,Index2,OverrideCycles,AdapterRead1,AdapterRead2`

NovaSeq X/X Plus flow cells contain 8 independent lanes. Illumina supports running different library preparation types in different lanes.

BaseSpace Sequence Hub (BSSH) Run Planning or the on-board built in run planning tool on the NovaSeq X/X Plus is recommended for setting up sample sheets for NovaSeq X/X Plus runs. **Run Planning** allows adding multiple configurations for runs and will **automatically populate** appropriate **OverrideCycles values** for Illumina library index adapter kits and user defined custom index adapter kits. When using multiple configurations, including a configuration with an index adapter kit with shorter index sequence lengths than the number of cycles specified for the read, an OverrideCycles value is automatically populated in the \[BCLConvert\_Data] section of the sample sheet and a base mask is applied to the cycles that are not part of the index sequence of the shorter index samples by default.

Run Planning allows users to select "Not Specified" for third party index adapter kits (and the index adapter kit does not need to be defined beforehand with this approach). When selecting "Not Specified" as the index adapter kit, Run Planning auto-populates an OverrideCycles value to use all cycles of the indexing reads by default for these samples; if the "Not Specified" sample index adapter sequences are shorter than the cycles used for the indexing reads, then the OverrideCycles value for these sample should be edited using rules below. It is possible for users to set up sample sheets manually without the use of Run Planning tools, if needed.

When running a sample with a **shorter index2 sequence length** than the specified cycles for index2, the **cycles at the beginning of index2 must be masked** in the OverrideCycles index2 value. Below is an example OverrideCycles value for a sample with dual 8 nucleotide index lengths for a run setup with 10 cycles for each indexing read (and 151 cycles for the sequencing reads).

OverrideCycles value: `Y151;I8N2;N2I8;Y151`

The NovaSeq X/X Plus v2 sample sheet settings guide highlights such an [OverrideCycles example](https://support-docs.illumina.com/SW/DRAGEN_v41/Content/SW/DRAGEN/SampleSheet.htm?Highlight=OverrideCycles).

In general, (for dual indexed libraries) when running different index length libraries on NovaSeq X/X Plus platform, below are the recommendations for setting OverrideCycles values for masking bases in shorter length index sequence libraries.

`IANB;NCID`

I: [Specifies an indexing read](https://support-docs.illumina.com/SW/DRAGEN_v41/Content/SW/DRAGEN/SampleSheet.htm?Highlight=OverrideCycles)

A: length of index1 sequence

N: Specifies cycles to be masked

B: (number of cycles specified for index1 read) - (length of index1 sequence)

C: (number of cycles specified for index2 read) - (length of index2 sequence)

D: length of index2 sequence

**A**,**B**,**C**,and **D** are all placeholders here and should be set to an appropriate integer.

The RunInfo.xml files produced by NovaSeq X/X Plus runs contains IsReverseComplement="Y" for the second index read, as below.

`<Read Number="3" NumCycles="10" IsIndexedRead="Y" IsReverseComplement="Y"/>`

Related Knowledge articles:

[How to add Custom Index Adapter and Library Prep kits to NovaSeq X/X Plus](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007789)

[How to use the 'OverrideCycles' setting in the Run Planning User Interface](https://knowledge.illumina.com/software/general/software-general-faq-list/000005671/?utm_source=agent\&utm_medium=email\&utm_campaign=kb)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8935), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008935%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/bclconvert-bcl2fastq/software-bclconvert-bcl2fastq-reference_material-list/000008935.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.