> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/tecan-iaps/microarray-tecan-iaps-reference_material-list/000002765.md).

# Tecan/IAPS with IAC Power Cycle Instructions

Performing a power cycle of the Tecan with Illumina Automation Control (Tecan with IAC), or the Infinium Automated Pipetting System with Illumina Automation Control (IAPS with IAC)--collectively referred to as the Tecan/IAPS system--can help maintain proper system performance and may resolve certain operational errors. The Tecan/IAPS with IAC system includes a personal computer (PC), Tecan/IAPS instrument, and water circulator. Follow these step-by-step instructions to power cycle the system in the correct sequence.

**Note:** This article is specifically for Tecan/IAPS with IAC. For Tecan/IAPS with ILASS, see [Tecan/IAPS with ILASS Power Cycle Instructions](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000008647.md).

**Why Is it important to Power Cycle the Tecan/Infinium Automated Pipetting System with Illumina Automation Control (Tecan/IAPS with IAC) Frequently?**

* Power cycling frequently helps alleviate pressure on and clear out the computer's RAM, which can become full and fragmented from running scripts over time. For high-throughput users, consider power cycling prior to each run.
* Power cycling allows the memory to unload completely and reduces the likelihood of the computer freezing under certain circumstances (ie, when using the IAC legacy software), which can cause the instrument to stop in the middle of runs.
* Proper power cycling resets all instrument memory and electronics which clears any temporary issues and allows everything to start fresh for the next run.
* As a troubleshooting tool, power cycling helps to resolve various anomalies, eg, instrument communication or initialization interruptions, LiHa (Liquid Handling) or RoMa (Robotic Manipulator) arms not functioning properly, water circulator communication issues, etc.

### **Power Cycle Recommendations for Tecan/IAPS with IAC**

* Illumina recommends power cycling the Tecan/IAPS system at least **once per week** at a minimum. Doing it more often is fine, only helpful, and not harmful.
* To prevent instrument connection issues, always exit IAC and/or power cycle the personal computer (PC) **first** before power cycling the Tecan/IAPS.
* To properly power cycle the Tecan/IAPS PC, the computer must be completely turned off and allowed to sit for at least 30 seconds, not simply restarted.
* After power cycling and initializing the instrument, IAC allows the robot to be initialized at any time thereafter via controls (**Sys Init**) within IAC. Power cycling the instrument may be necessary at times if IAC fails to initialize the instrument.
* It is recommended to turn off the Tecan/IAPS with IAC instrument and PC at the end of each day, or after use to save energy and help preserve the electronics. However, it is not necessary, and the instrument can be left on between periodic power cycles.

### **Water Circulator Recommendations for Tecan/IAPS with IAC**

* It is critical to leave the water circulator in Standby mode to avoid initialization errors, even if it is not being actively used.
* The water circulator should only be turned off for an extended period of time using the switch on the back of the unit. If the PC and instrument are also turned off **at the same time**(eg, in between uses), then turn everything back on when the system is used again.
* In general, it is only necessary to power cycle the water circulator by itself as a troubleshooting step, with the PC and instrument left on, and only **before or after** the instrument is initialized, but not during initialization, to help reduce communication issues. If any issues arise, ensure the circulator is on and run **Sys Init** again to re-initialize the entire system.
* The water circulator can always be power cycled by itself with the PC and instrument on and IAC launched (open). Avoid power cycling the water circulator during the Tecan/IAPS initialization process. If any issues arise, ensure the circulator is on and run **Sys Init** again to re-initialize the entire system.

### **Step by Step Instructions for Tecan/IAPS with IAC Power Cycle**

1. Exit and close IAC, if running, first.
2. Shut down the computer and wait at least 30 seconds.
3. Turn off the Tecan/IAPS instrument for at least 30 seconds.
4. Turn off the water circulator at this point using the main switch on the lower back panel of the circulator. Wait 10 seconds, then turn it back on. This can also be done independently with both the PC and instrument on and connected, as indicated above. In either case, the water circulator will display an hourglass while booting up. Ensure it eventually boots into Standby mode.
5. Turn on the computer.
6. Turn on the Tecan/IAPS instrument.
7. Enter the user name and password for the instrument control computer.
8. After the "Standby" screen appears on the water circulator control panel, launch the IAC control software.
   1. Select **Yes** to initialize the robot when the prompt appears.
   2. If necessary, check and activate the water circulator through the IAC software to verify a connection to the controlling computer.
      1. Double check the circulator is online by selecting **File > Configuration**… from the top menu to ensure that the General Circulator Property **CirculatorComPort** is assigned and that **CirculatorOperatingMode** says **Online**.
      2. Next, select **Robot QC Tasks** from the workflow menu on the left.
      3. Select **Circulator Manager**.
      4. Use the drop-down menu to select **Circulator** - **Activate**.
      5. Select **Execute**. Confirm the circulator is now on and therefore connected to the computer.
      6. Use the drop-down menu to select **Circulator** - **Standby**, then select **Execute** to put the circulator back into Standby mode so it is ready for use.

**References:**

* [Infinium Lab Setup and Best Practices](https://support.illumina.com/downloads/infinium-assay-lab-setup-and-procedures-11322460.html) > [Robot (Infinium Automated Pipetting System) Usage](https://support-docs.illumina.com/ARR/InfLabSetup/Content/ARR/SetupandProcedures/RobotIntro_fINF.htm)
* [Infinium LiHa and Infinium RoMa Operating Manual](https://support.illumina.com/downloads/infinium-liha-infinium-roma-manual-1000000110155.html)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2765), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002765%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/tecan-iaps/microarray-tecan-iaps-reference_material-list/000002765.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.