> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/tecan-iaps/microarray-tecan-iaps-faq-list/000009625.md).

# What Tecan/Illumina instrument and software combinations require the use of Tip Guides?

The Tecan/Infinium Automated Pipetting System (IAPS) with Illumina Lab Automation Software Solution (**ILASS**) is used to perform automated Infinium workflow for Infinium EX Genotyping and EX Methylation formats, as well as selected legacy Infinium formats (HD Methylation, LCG-8, LCG-PGx). The Tecan/IAPS with ILASS uses Tecan fixed tips and does not use tip guides for BeadChip sample hybridization.

The Tecan/Infinium Automated Pipetting System (IAPS) with Illumina Automation Control (**IAC**) software is used to perform automated Infinium workflow for legacy Infinium BeadChip formats such as Infinium HD, HTS, LCG, and XT. The Tecan/IAPS with IAC uses Illumina Teflon tips and requires tip guides that are placed on top of the BeadChip to dispense samples properly during the Hybridization step.

**Table1.** Overview of the compatible BeadChip formats, LiHa Tip types and Tip Guide requirement on both systems.

| **Instrument and software** | **Compatible BeadChip formats**                                                       | **LiHa Tip type**    | **Tip guide** |
| --------------------------- | ------------------------------------------------------------------------------------- | -------------------- | ------------- |
| Tecan/IAPS with **IAC**     | HD (Ultra, Super, Methylation), HTS, LCG (8-sample, Quad), XT                         | Illumina Teflon tips | Required      |
| Tecan/IAPS with **ILASS**   | EX-24 and EX-48 (Genotyping, PGx, Methylation), HD Methylation, LCG\* (8-sample, PGx) | Tecan fixed tips     | Not used      |

\* LCG-8 samples enabled on ILASS v1.6 only (Note: ILASS v1.6 does not support Clarity LIMS integration). A Quick Start Protocol for LCG semi-automated workflow is available for ILASS v1.6. Contact <techsupport@illumina.com> for more information.

For additional information, see the following:

* [Infinium LiHa and Infinium RoMA Operating Manual](https://support.illumina.com/downloads/infinium-liha-infinium-roma-manual-1000000110155.html)
* [Infinium RoMA (IAPS with ILASS) Product Documentation](https://support.illumina.com/downloads/infinium-roma-product-documentation.html)
* [Illumina Automation Control Release Notes](https://support.illumina.com/downloads/illumina-automation-control-release-notes.html)
* [ILASS Infinium Module Release Notes](https://support.illumina.com/downloads/ilass-release-notes.html)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9625), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009625%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/tecan-iaps/microarray-tecan-iaps-faq-list/000009625.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.