> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/non-human-methylation/microarray-non-human-methylation-reference_material-list/000001566.md).

# Infinium HD Methylation assay workflow for fresh/frozen versus FFPE samples

All DNA samples used with the Infinium HD Methylation assay, whether fresh/frozen or formalin-fixed, paraffin-embedded (FFPE), require a Zymo bisulfite conversion kit (certain Zymo kits have a modified Illumina protocol).

FFPE samples also require the [Infinium FFPE QC Assay and Infinium HD FFPE DNA Restore kits](https://www.illumina.com/products/by-type/molecular-biology-reagents/infinium-ffpe-qc-dna-restoration.html). This article describes the differences in workflow for fresh/frozen versus FFPE DNA samples.

![](/files/62dJgX2fhkaGwbsOcrNd)

**Figure 1:** Graphical overview of the Infinium HD Methylation assay for fresh/frozen versus FFPE DNA samples.

1. Isolate DNA using a kit of your choice, then quantify DNA using a **fluorometric method** (eg, PicoGreen, Qubit) to ensure accurate measurement of double-stranded DNA.

**Note:** Accurate DNA quantification at this step is critical. Following bisulfite conversion, all subsequent steps deal in volumes only and rely on the assumption that DNA input was accurately measured at this point.

2. Evaluate sample quality (**FFPE only**).For human samples, check DNA quality using the [Infinium FFPE QC kit](https://www.illumina.com/products/by-type/molecular-biology-reagents/infinium-ffpe-qc-dna-restoration.html) (see [Support page](https://support.illumina.com/array/array_kits/infinium_ffpe_qc.html)) and a qPCR kit with green fluorescent dye of your choice. Illumina recommends proceeding only with samples with **delta Ct (or delta Cq) < 5** (refer to the [QC protocol](https://support.illumina.com/downloads/infinium_hd_ffpe_qc_assay_insert_\(15020981_rev_c\).html)).
3. Perform **bisulfite conversion (all sample types).**

Recommended DNA input for Zymo Bisulfite Conversion kit (refer to the approved kits in Table 2):

**Fresh/frozen:** > 250 ng (manual) or 1000 ng (automated).\
**FFPE:** ≥ 250 ng (up to 1000 ng).

**Note:** Using > 250 ng of DNA, such as 500 ng, is recommended for the best results and reproducibility, especially for FFPE samples. When using Lightning MagPrep kits, the recommended FFPE DNA input is 1000 ng.

**Table 1. C**omparison of Zymo classic vs Lightning MagPrep bisulfite conversion kits.The classic kits follow a modified protocol.

|                             | **Zymo D5004, D5001, or D5002 (classic kits)**                                                                                                                                                                                                                                                                                                                                                                      | **Zymo D5046, D5047 or D5049 (Lightning MagPrep kits)**                                            |
| --------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | -------------------------------------------------------------------------------------------------- |
| **Thermal cycler protocol** | <p>Follow the Zymo bisulfite conversion protocol with the following <em>critical</em> adjustment to the thermal cycler program (end of Day 1), as stated in the Zymo protocol appendix:</p><p><br>Incubate samples as follows for 16 cycles:</p><ul><li>95°C for 30 seconds</li><li>50°C for 1 hour</li></ul><p>Then, hold at 4°C until ready to proceed with the Zymo bisulfite conversion protocol for Day 2.</p> | Follow the Zymo bisulfite conversion protocol with no modifications to the thermal cycler program. |
| **Elution**                 | Elute bisulfite-converted samples (Fresh/frozen and FFPE) in 12 µl (manual) or 22 µl (automated) buffer provided by Zymo.                                                                                                                                                                                                                                                                                           | Elute bisulfite-converted samples (Fresh/frozen and FFPE) in 25 µl buffer provided by Zymo.        |

4. Restore DNA samples (**FFPE only**). Use the entire bisulfite-converted eluate (\~**8 µl**) as input for the [Infinium HD FFPE DNA Restore kit](https://support.illumina.com/array/array_kits/infinium_ffpe_dna_restoration_solution.html), and follow the [restore protocol](https://support.illumina.com/downloads/infinium_hd_dna_restoration_protocol_\(15014614_rev_c\).html).
5. Proceed with the Infinium Methylation Assay Amplify DNA step.

**Fresh/frozen:** Use **4 µl** bisulfite-converted DNA input for the [Infinium HD Methylation Assay](https://support.illumina.com/downloads/infinium-hd-methylation-reference-guide-15019519.html) protocol, beginning with the steps to make the **MSA4** plate.

**FFPE:** Use **8 µl** restore kit eluate as input in the [Infinium HD FFPE Assay](https://support.illumina.com/downloads/infinium-ffpe-assay-reference-guide-15021525.html) with the steps to make the **MSA5** plate.

6. After completing the Resuspend DNA step, proceed to Hybridize DNA to the BeadChip.

* Manual protocol.
  * For MethylationEPIC (8x1 BeadChips), make sure to Hyb **26 µl**, rather than 15 µl.
  * For Mouse Methylation (12x1 BeadChips), Hyb **15 µl**.
* Automated protocol using Tecan/IAPS with IAC: Select **MSA4 Tasks** in Illumina Automation Control (IAC) software and proceed to Hyb Multi-BC2.
  * For MethylationEPIC BeadChip, select the 8-sample BeadChip picture and use Tip Guide-E for Hybridization.
  * For Mouse Methylation arrays, select the leftmost 12-sample BeadChip picture and use Tip Guide-B for Hybridization.
  * Refer to the [Infinium HD Methylation Assay Reference Guide](https://support.illumina.com/downloads/infinium-hd-methylation-reference-guide-15019519.html) for more information.

**Table 2.** Additional Kits Required for Infinium HD Methylation:\
Illumina has validated and recommends only the [Zymo Research](https://www.zymoresearch.com/) Bisulfite Conversion kits listed in this table. All kits listed are compatible with fresh/frozen and FFPE DNA samples.\
![](/files/ZzsGiaWf9jtEn9BLo0FO)\
\&#xNAN;**\***&#x49;f using the Lightning MagPrep kits, the recommended FFPE DNA input is 1000 ng.\
**For additional information,** refer to:\* [Infinium MethylationEPIC BeadChip Support Resources Page](https://support.illumina.com/array/array_kits/infinium-methylationepic-beadchip-kit.html)

* [Infinium Mouse Methylation BeadChip Support Resources Page](https://support.illumina.com/array/array_kits/infinium-mouse-methylation-beadchip-kit.html)
* Illumina Knowledge Articles:
  * [Which bisulfite conversion kit should be used for the Infinium Methylation arrays?](/microarray/general/microarray-general-reference_material-list/000001524.md)
  * [DNA Input Guidelines for Infinium MethylationEPIC Arrays](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/DNA-Input-Guidelines-for-Infinium-MethylationEPIC-Arrays/README.md)
* Methylation Recorded Webinars
  * [Infinium Methylation: Introduction to GenomeStudio analysis and BeadArray Controls Reporter](https://www.youtube.com/embed/tuY7DEMSolE?autoplay=1\&rel=0)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1566), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002679%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/microarray/non-human-methylation/microarray-non-human-methylation-reference_material-list/000001566.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.