> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/human-methylation/microarray-human-methylation-reference_material-list/000008604.md).

# Infinium EX Methylation DNA input guidelines and compatible bisulfite conversion kits

### **Supported Sample Types**

* The Methylation Screening Array (MSA) is validated for blood samples.
* MSA is **not validated** with FFPE or saliva samples.

### **Recommended DNA Input**

* The recommended DNA input for the bisulfite conversion reaction (manual) is minimum 50 ng, and up to 1 μg.
* For optimal results and higher reproducibility, use higher DNA input.

Refer to the [Infinium EX Methylation Assay Product Documentation](https://support.illumina.com/downloads/infinium-ex-methylation-assay-product-documentation.html) for additional information.

**Supported Bisulfite Conversion Kits**

Before carrying out the Infinium EX Methylation Assay, samples must first be bisulfite converted. Illumina has validated and recommends only the following kits from [Zymo Research](http://www.zymoresearch.com/) for Infinium EX Methylation (Table 1).

**Table 1.** EZ-96 DNA Methylation Lightning MagPrep Kits (uses magnetic beads for purification). Bisulfite conversion completes in \~ 3 hours.

| **Product Name**                        | **Size**    | **Format**     | **Catalog No.** | **Elution Volume** | **Preparation** |
| --------------------------------------- | ----------- | -------------- | --------------- | ------------------ | --------------- |
| EZ-96 DNA Methylation-Lightning MagPrep | 4 x 96 Rxns | 96-well (Deep) | D5046           | 25 μl              | Manual          |
| EZ-96 DNA Methylation-Lightning MagPrep | 8 x 96 Rxns | 96-well (Deep) | D5047           | 25 μl              | Manual          |

**Note:** Zymo EZ DNA Methylation kits (D5001, D5002, D5004) have **not** been validated and are not supported for use with Infinium EX Methylation.

**Recommended DNA Sample volumes for Zymo Bisulfite Conversion**

**For Zymo kits D5046 and D5047, in the first step of Bisulfite Conversion of the Zymo protocol, it states:**

Add 130 µl of Lightning Conversion Reagent to 20 µl of a DNA sample in a Conversion Plate. Mix the samples by pipetting up and down. **Note:** If the volume of DNA is less than 20 µl, compensate with water.

**Note:** DNA quantification must be performed **before** bisulfite conversion. After bisulfite conversion, the DNA does not need to be quantified and the user can proceed directly to the Infinium EX Methylation Assay.

**Additional information:**

* For information on how the Lightning kits were validated, refer to the Illumina Technical Note: [Automated bisulfite conversion for Infinium Methylation BeadChips](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/automated-bisulfite-infinium-methylation-tech-note-m-gl-00144/automated-bisulfite-Infinium-methylation-tech-note-m-gl-00144.pdf)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8604), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008604%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/human-methylation/microarray-human-methylation-reference_material-list/000008604.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.