Evaluation of Infinium Staining Controls in the GenomeStudio Controls Dashboard
All Infinium BeadChips have built-in control probes designed to assess array and sample performance. While individual controls can be informative of specific steps in the Infinium workflow, they may be affected by other steps in the workflow. Therefore, it is recommended to assess the built-in Infinium controls in combination with each other.
Considerations for evaluating Infinium controls
Infinium controls are evaluated based on relative intensities rather than absolute intensity thresholds. Signal intensities in the Infinium assay are subject to variations that are rooted in, but not limited to:
Variability in DNA preparation methods, sources, or tissue types
Operator variability
Variability between scanners that do not impact results
Inconsistencies in reagent storage temperature
These factors can manifest as batch-to-batch variability. Process controls are used to ensure that reagent, BeadChip, and scanner performance are within specification.
Sample-Independent controls evaluate BeadChip and reagent performance, DNA hybridization efficiency, and the staining process (XStain) of the Infinium workflow. Staining Control is one of the Sample-Independent controls, along with Extension, Target Removal, Hybridization, and Restoration Controls.
Staining Control Chemistry
Staining Controls consist of beads covered with either high or low (background) levels of oligonucleotides bound to dinitrophenyl (DNP) or biotin (Fig. 1).

Figure 1. Staining Control beads bound to DNP or biotin
On the BeadChip, the DNP on the control beads is directly stained with red fluorescence-labeled anti-DNP antibodies, which are counterstained with DNP-conjugated secondary antibodies. Similarly, biotin is directly stained with green fluorescence-labeled streptavidin, which is counterstained with biotinylated anti-streptavidin antibodies (Fig. 2). Three rounds of staining and two rounds of anti-staining during XStain increases fluorescent signal intensities. The Staining Controls are then monitored in their respective channels, red (DNP) or green (biotin).

Figure 2. Fluorescence signal amplification of Staining Controls
Expected Staining Control Outcomes
Staining Control beads are manufactured with short oligonucleotides labeled with either DNP or biotin, while other controls are made with standard oligo-bound beads. Due to this distinct chemistry, Staining Controls perform differently than the other controls. Because DNP and biotin are directly attached to the oligos, Staining Controls are independent of both DNA hybridization to the probe and single-base extension.
Table 1. Expected Staining Control outcomes
Purpose
Name
Evaluate Green
Evaluate Red
Expected Intensity
Staining Red
DNP (High)
-
+
High
Staining Red
DNP (Bgnd)
-
+
Low
Staining Green
Biotin (High)
+
-
High
Staining Green
Biotin (Bgnd)
+
-
Low
For the Staining Red controls, strong positive signals are expected in the red channel for the DNP (High) data points. The DNP (Bgnd), Biotin (High), and Biotin (Bgnd) signals are expected at background levels in the red channel.
For the Staining Green controls, strong positive signals are expected in the green channel for the Biotin (High) data points. The Biotin (Bgnd), DNP (High), and DNP (Bgnd) signals are expected at background levels in the green channel.
No comparison is made between the levels of positive signal in the red and green channels, nor is there a specific threshold value for expected intensities.

Figure 3. Expected Staining Controls profile in the GenomeStudio Controls Dashboard.
Troubleshooting
While all controls rely on successful staining, low Staining Control intensities do not necessarily indicate that the sample(s) failed. If the other controls that rely on successful staining are as expected and the sample metrics* meet array specifications, low Staining Control intensities are not expected to impact data quality of the responding samples. Illumina recommends evaluating the controls in the Controls Dashboard in combination with each other, along with evaluation of the sample metrics, for a complete assessment of array and sample performance.
*Metrics include Call Rate and Log R Dev scores for genotyping arrays, Detected CpG values for methylation arrays.
References:
For any feedback or questions regarding this article (Illumina Knowledge Article #6555), contact Illumina Technical Support techsupport@illumina.com.