> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/genomestudio/microarray-genomestudio-reference_material-list/000002624.md).

# Information on cnvi-labelled (Intensity Only) probes in Infinium Arrays

The cnvi-labelled probes correspond to, and are synonymous with, Intensity Only markers. These probes were added to earlier BeadArray designs to provide better genomic coverage and to support Copy Number Variation (CNV) analysis for detecting copy number changes.

These probes target regions where no SNPs were known at the time the array content was developed. They can also fill gaps between polymorphic sites, provide coverage in regions underrepresented by SNPs, or replace SNPs that do not perform well.

Intensity Only probes are designated as out-of-stats markers and are used only for intensity information in copy number calculations. In contrast, in-stats markers are SNPs used to calculate SNP statistics, such as call rate, heritability, and reproducibility.

Intensity Only probes can also be used in cases where only one colour is detected and the intensity of that colour is measured. This can occur with transversion SNPs, such as \[A/T] and \[C/G], where Infinium I chemistry is used. Infinium I chemistry uses two allele-specific probes on the array, with each probe designed to detect one allele of the transversion SNP. In this case, the intensity rather than the specific colour is measured.

Intensity Only probes can also be found in IN/DEL SNPs, where a single probe is designed to detect either the inserted or deleted allele and a two-colour measurement is not required.

In the Manifest File, Intensity Only markers are designated as \[N/A] in the SNP column and assigned a value of 1 in the Intensity Only column. These probes are generally identified by the prefix cnvi, indicating that they are Intensity Only probes rather than standard SNP probes on the array.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6b39933bda4693a446e2b0b395dc6bdf10a59c64%2Fimage1.png?alt=media)

**Figure 1**. cnvi probes in the manifest CSV file, showing \[N/A] in the SNP column and a value of 1 in the Intensity Only column.

The cnvi or Intensity Only probes are located in regions with high copy number variation (CNV) and are not intended for SNP calling. In general, regions with high CNV do not contain SNP probes because SNP probes in these regions tend to not perform well. Because cnvi probes target non-polymorphic sites rather than SNPs, no SNP data are associated with these probes.

These probes provide intensity information only and do not contain genotype information. As a result, they generate No Calls (NC) in the GenomeStudio Full Data Table.

Because cnvi probes are not used for genotype calling, they can be filtered out when reviewing genotype data. GenomeStudio does not include these loci when calculating the call rate.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8427cda4bcdaca4592ce2f43852b5e88a7d50e1c%2Fimage2.png?alt=media)

**Figure 2**. cnvi probes in the GenomeStudio Full Data Table showing No Calls (NC) in the GType column.

Typical examples of cnvi probes displayed in the Genome Studio SNP Graph are illustrated as follows.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-246550ea284dc57909617ff7ab5aff4ee2e05a28%2Fimage3.png?alt=media)\
![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ed3d61f8d5c159beeac06cec1816f1e7e76656bc%2Fimage4.png?alt=media)

**Figure 3**. Examples of cnvi probes in the GenomeStudio SNP graph, showing the intensity distribution of samples for each probe.

With the discovery of additional SNPs and the development of higher-density arrays with broader genomic coverage, the need for cnvi probes has decreased over time.

**Reference**:

[Interpreting Infinium Assay Data for Whole-Genome Structural Variation](https://www.illumina.com/Documents/products/technotes/technote_cytoanalysis.pdf)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2624), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002624%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/genomestudio/microarray-genomestudio-reference_material-list/000002624.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.