> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-troubleshooting-list/000009270.md).

# Recommendations for Infinium Methylation Screening Array Quality Control Analysis

### **Which software is recommended for Methylation Screening Array Data QC?**

Illumina offers multiple software tools for Infinium Methylation Array quality control (QC) analysis. QC assessment can be performed using GenomeStudio 2011.1 Methylation Module or DRAGEN Array Methylation QC, or a combination of both. For a comparison of methylation analysis software features, refer to the [Methylation Array Data Analysis Tips](https://www.illumina.com/techniques/microarrays/methylation-arrays/methylation-array-data-analysis-tips.html) page.

[DRAGEN Array Methylation QC,](https://help.dragenarray.illumina.com/overview/our-features#dragen-array-methylation-qc) a cloud-based application, is the recommended method for quality control analysis of Methylation Screening Array (MSA) data. DRAGEN Array Methylation QC delivers high-throughput and quantitative evaluation of Infinium Methylation Arrays control metrics. It also includes control plot and CG results. For additional information on the sample QC methods and metrics employed by the software, review the Technical Note [High-throughput Infinium methylation array QC using DRAGEN Array Methylation QC software](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/dragen-array-methylation-qc-tech-note-m-gl-02644/dragen-array-methylation-qc-tech-note-m-gl-02644.pdf).

### **Why would a user still use GenomeStudio and/or BeadArray Controls Reporter?**

Users may choose to continue using GenomeStudio Methylation Module and/or BeadArray Controls Reporter (BACR) for several reasons:

* Control probe performance is easier to visualize and review in GenomeStudio, while flagged controls are clearly highlighted in the BACR output file.
* The Probe Detection and Reproducibility specifications listed in the [Methylation Screening Array Data Sheet](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/infinium-methylation-screening-array-data-sheet-m-gl-01893/infinium-methylation-screening-array-data-sheet-m-gl-01893.pdf) (see also Table 1) are based on analysis performed using the GenomeStudio Methylation Module.
  * The 96% Detected CpG specification applies to GenomeStudio analysis, whereas, DRAGEN Array does not use the probe detection rate as a criterion to determine sample pass/fail status. For a comparison of the probe detection calculations between the two applications, refer to the DRAGEN Array User Guide under section [DRAGEN Array Methylation QC and GenomeStudio Methylation Module Differences](https://help.dragenarray.illumina.com/product-guides/dragen-array-cloud-analysis/overview/dragen-array-methylation-qc#dragen-array-methylation-qc-and-genomestudio-methylation-module-differences).
  * Sample reproducibility can be evaluated in GenomeStudio but is not currently assessed in DRAGEN Array.
* BACR control thresholds can be adjusted according to the recommendations provided in the DRAGEN Array User Guide under section [Methylation QC Threshold Adjustment](https://help.dragenarray.illumina.com/product-guides/dragen-array-cloud-analysis/overview/dragen-array-methylation-qc#section-methylation-qc-threshold-adjustment).
* DRAGEN Array Methylation QC is a cloud-based application. Users who prefer to review their data using local analysis tools can continue to use GenomeStudio and/or BACR.

**To complement GenomeStudio/BACR data QC results with DRAGEN Array Methylation QC:**

* Review the [sample\_qc\_summary file](https://help.dragenarray.illumina.com/product-guides/output-files#methyl_qc_report) generated in DRAGEN Array. This file provides the **passQC** and **prop\_probes\_passing** metrics, which indicate whether a sample passed or failed the DRAGEN Array QC metrics.
* The probe detection rate is typically lower in DRAGEN Array than in GenomeStudio Methylation Module because DRAGEN Array uses a more stringent detection p-value calculation.
* Differences between DRAGEN Array and GenomeStudio algorithm are expected and are described in the Technical Note [High-throughput Infinium methylation array QC using DRAGEN Array Methylation QC software.](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/dragen-array-methylation-qc-tech-note-m-gl-02644/dragen-array-methylation-qc-tech-note-m-gl-02644.pdf)

###

Both, GenomeStudio and DRAGEN Array are required to obtain the performance metrics described in the Methylation Screening Array documentation.

**Table 1.** Summary of Methylation Screening Array performance specifications and the software used to generate or evaluate each metric.

| **Specification**        | **Software**                    |
| ------------------------ | ------------------------------- |
| Detected CpG >96%        | GenomeStudio Methylation Module |
| Reproducibility r ≥ 0.98 | GenomeStudio Methylation Module |
| Passing QC               | DRAGEN Array Methylation QC     |

**References:**

[Introduction to GenomeStudio and BeadArray Controls Reporter for Infinium Methylation Support Webinar](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000008062.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9270), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009270%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-troubleshooting-list/000009270.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.