> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-troubleshooting-list/000005545.md).

# How to switch between AutoLoader Mode and Manual Mode when using iScan

When operating an Illumina iScan with an AutoLoader 2.x, users might need to switch between AutoLoader 2.x and Manual mode to use the iScan with or without the AutoLoader 2.x.

* To run the iScan System without the AutoLoader 2.x, switch the reader into Manual mode.
* To return to using AutoLoader 2.x, simply switch the readers back into AutoLoader mode.

Switching from AutoLoader 2.x Mode to Manual Mode

1. Complete all automated scans.
2. Remove all carriers from the AutoLoader 2.x input stacks.
3. In the iScan Control Software (ICS), select the Menu button and select 'Exit'.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-27ba8c6ce8f5a224307aa09c04fed98948198af3%2Fimage1.png?alt=media)

**Figure 1.** ICS menu with the 'Exit' button highlighted.

4. Restart the iScan Control Software. The software then prompts the user to run the iScan system in AutoLoader mode or Manual mode.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d592d0d2c09badcfe92589e5509b9fb989972bfd%2Fimage2.png?alt=media)

**Figure 2.** ICS Scan Mode selection upon restarting the software.

5. Select 'No' to run the iScan System in Manual mode. For information on manual scanning, refer to the [iScan System Guide](https://support-docs.illumina.com/ARR/iScan/Content/ARR/FrontPages/iscan.htm).

Switching from Manual Mode to AutoLoader 2.x Mode

1. Complete all manual scans.
2. Place the carriers to be scanned on the AutoLoader 2.x input stacks.
3. To run the AutoLoader 2.x in dual-reader mode, make sure the iScan Control Software is closed on the remote reader.
4. Follow the instructions for scanning BeadChips using an AutoLoader 2.x. Refer to section on Scan BeadChips with AutoLoader 2.x in the [AutoLoader 2.x Product Documentation](https://support.illumina.com/downloads/autoloader_2x_user_guide_\(15015394_a\).html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5545), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005545%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-troubleshooting-list/000005545.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.