> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001492.md).

# Port Reset and Power Cycle Sequence for iScan or HiScan, without an AutoLoader

A complete power cycle of the instrument can help resolve many communication issues. For iScan/HiScan systems, this involves turning the instrument off and back on, as well as resetting the communication ports in the correct sequence. Follow the complete power cycle procedure outlined below to ensure optimal results.

**Before starting:**

* Make sure that the computer's local (C:) hard drive has free memory space and is not full.
* Check that the iScan is plugged in with no loose cables.
* Try restarting the iScan computer by selecting **Start** > **Restart**, instead of Shut Down, then attempt to start the iScan Control Software (ICS).

**Use Device Manager to clear the ports, then power cycle the instrument and computer in the following order:**

1. Close ICS and leave the iScan on. If necessary, use Task Manager to force-quit ICS and other open programs.
2. Launch **Device Manager:** Using the Windows Search or Control Panel Search, type 'Device Manager' and select the program to open.
3. In Device Manager, navigate to **Ports (COM & LPT)**.
4. The number of ports listed in Device Manager will depend on the model of the iScan computer (Table 1): **Table 1.** **List of expected number of ports for each iScan computer model**\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-bdd50f86d5501428204903b55e6428f08d1e5808%2Fimage1.png?alt=media)
5. Starting at the bottom of the port list, right-click on the port, and select 'Uninstall Device'. For example, in Figure 1, start with (COM9). If prompted, check the 'Delete the drivers software for this device' option then select 'OK'. ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-9b2befbac9d4b35ca7930b6bb28be88f7713d0aa%2Fimage2.jpg?alt=media)

**Figure 1**: **Example of 18 ports listed in Device Manager for a DC Fortris A8202 computer**

6. Repeat this process to uninstall all COM ports except for Communication Port (COM1) and Phoenix Serial Port (typically COM 3 or COM7, Figure 2):

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-5c41e5de327bbac0d01a81ae76b9fd3307f075ce%2Fimage3.jpg?alt=media)

**Figure 2**: **All ports removed, except for COM1 and Phoenix Serial Port.**

7. Turn the scanner **OFF** using the hard switch on the back of the instrument.
8. Sometimes ports will re-populate when the scanner is shut off (Figure 3). Uninstall remaining ports in Device Manager.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8aaab42d548a57d2682cf5b2f1f00e8175d7d204%2Fimage4.jpg?alt=media)

**Figure 3: Example of ports re-populating after iScan is turned off**

9. Turn the computer **OFF** by selecting **Shut Down** from the Start menu and **wait 3 minutes**.
10. Restart the computer and let the computer fully reboot (\~2 minutes).
11. When the computer is fully booted, turn the scanner **ON**.
12. In File Explorer, open **My Computer** or **This PC** and confirm that the iScan 'Archimedes'/'Do Not Eject' drive is detected. This drive must be recognized by the computer before opening ICS.
13. Open Device Manager and confirm that the expected amount of ports are populated, and that there are no duplicate or missing ports.
14. Start ICS and scan the BeadChips.

Reference for starting or shutting down the iScan: [iScan System Product Documentation](https://support.illumina.com/downloads/iscan-system-product-documentation.html)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1492), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001492%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001492.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.