> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-faq-list/000005924.md).

# Can Infinium BeadChips that have been frozen still be used?

Infinium BeadChips that have been exposed to freezing conditions should not be used, as freezing can compromise both their physical integrity and assay performance.

The beads on the surface of the BeadChip are held within microscopic wells (pits) by relatively weak Van der Waals forces. When BeadChips are subjected to freezing temperatures and subsequent thawing, the materials of the BeadChip undergo expansion and contraction. These temperature-induced stresses can cause beads to become dislodged from the wells, resulting in bead loss and reduced data quality or assay failure.

Freezing may also damage the chemical integrity of the beads. The bead surfaces are functionalized with oligonucleotides that are essential for hybridization and signal detection. Exposure to freezing conditions can alter the bead surface or cause the oligonucleotides to shear off, leading to reduced signal intensity, poor reproducibility, and unreliable results.

BeadChips are shipped at room temperature. Upon receipt, they should be stored at 2-8°C and should not be frozen under any circumstances. For complete guidance on proper storage conditions, refer to the article [The correct storage temperature for Infinium BeadChips is 2-8°C](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000002839.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5924), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005924%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-faq-list/000005924.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.