Guidelines for cleaning the Te-Flow
The Te-Flow Module is an incubation apparatus used in the Infinium Assay workflow to process up to 48 BeadChips assembled in individual flow-through chambers for XStain.
Always switch off the instrument before cleaning the surface of Te-Flow Module components located on the instrument worktable.
Use caution when working with cleaning agents, as they may be hazardous. Always follow the safety precautions provided by the manufacturer. Strong detergents can damage the surface coatings of the carrier and worktable. Use alcohol or water to clean the instrument.
Refer to the following tables for the recommended cleaning agents for each component of the Te-Flow.
Table 1. Recommended cleaning agents for Te-Flow module components.
Commercially Available Cleaning Agents
Module Component
Subcomponent
Cleaning Agent
Flow-through chamber
Frame
Alcohol, water
Clamps
Alcohol, water
Glass back plate
Water, Alconox, bleach
Spacer
Dispose
Chamber rack
Heat exchanger
Alcohol, water
Reagent pan
Alcohol, water
Waste tubing leading to waste carboy
Bleach*
*If needed, bleach can be flushed through the waste tubing leading to the waste carboy to clean the lines.
Table 2. Cleaning agents specifications.
Cleaning Agents Specifications
Agent
Specification
Water
Use distilled or deionized water
Alcohol
Use 70% ethanol or isopropanol (2-Propanol)
To prevent damage to the reagent pan, avoid cleaning with the following chemicals:
Aromatic carbohydrates
Oxidizing acids
Perchlorethylene
Cleaning the Surface of the Chamber Rack
Make sure that the surface of the chamber rack heat exchanger is clean.
If necessary, clean the surface using Kimwipes dampened with alcohol or laboratory-grade water
Dry the surface with clean, dry Kimwipes.
Dispose of used Kimwipes with the formamide waste.

Figure 1. Heat exchanger of the chamber rack (from the rear left).
Cleaning the Reagent Pan
To clean the reagent pan, proceed as follows:
Remove the chamber rack from the reagent pan and place the top part of the chamber rack, ie, the heat exchanger, on the worktable.
Tighten the waste tubing of the reagent pan with the hose clamp at all three ends of the y-connector and remove the waste tubings leading to the chamber racks from the y-connector.
Rinse the reagent pan with tap water to remove remaining staining reagents.
Let the reagent pan dry, reassemble the chamber rack and connect the waste tubing to the y-connector.

Figure 2: Reagent pan of the chamber rack (from the right back). A: Outlet to waste pump
Warning: The contents of the reagent pan, waste tubing, and reagent waste container are potentially hazardous. Wear gloves and safety goggles when handling these Te-Flow Module components. Dispose of reagent waste and/or replaced parts in accordance with all applicable local, state, and federal environmental, health, and safety laws and regulations.
Emptying the Reagent Waste Container
After each operation, the liquid level in the reagent waste container must be checked and, if necessary, the container must be emptied, as follows:
Tighten the waste tubing at the inlet of the reagent waste container using a hose clamp.

Figure 3. Tightening a tubing with a hose clamp. 2. Pull the waste tubing out of the reagent waste container. 3. Empty the reagent waste container. 4. Place the reagent waste container back to the left side underneath the instrument. 5. Push the end of the waste tubing back into the reagent waste container. 6. Make sure that the waste tubing are not forming loops in the reagent waste container (or anywhere else). Warning: Make sure that the waste tubing does not form loops. Loops may cause reagent backflow, which can result in overflow from the reagent pan onto the worktable. 7. Loosen the screw of the hose clamp. Warning: The liquid in the waste tubing and reagent waste container is potentially hazardous. Wear gloves and safety goggles when handling these components. Dispose of reagent waste in accordance with all applicable local, state, and federal environmental, health, and safety laws and regulations.
References:
For any feedback or questions regarding this article (Illumina Knowledge Article #5670), contact Illumina Technical Support techsupport@illumina.com.