# Resetting the TeFlow Water Re Circulator Air Filter Alarm

The VWR water re-circulator may display these messages with information regarding maintenance:

"Service air filter and reset"

![](/files/wY8GR88pu6jMkCnuel5D)

"The fluid should be replaced"

![](/files/96JvpLnLRIlyklfpw5hZ)

**Follow these instructions to reset the alarm:**

1. Make sure the water circulator is off or in standby mode.

![](/files/GS2kZ8fP3VNQxnABdn7r)

2. Open the Front Panel by pulling forward. The air filter can now be removed and brushed clean if there is excessive dust. Reseat the filter and the front panel.

![](/files/4ZNpJHRFIUwQW7VPFJmx)

3. To reset the counter, turn on the re-circulator and press the Menu button (see Fig 3).

![](/files/ur133ANaZKmQ0kO1all0)

4. Press the Service Button on the touch screen to access the Service Menu.
5. Press both the "Reset" buttons on the right hand side next to "Air Filter Service Interval" and "Fluid Service Interval" (See Fig 4). This will reset the counter to 180 days.

![](/files/VeM9pS22iHLp973vSOIe)

For additional information, see [VWR Circulating Baths User Manual](https://digitalassets.avantorsciences.com/adaptivemedia/rendition?id=4add559d18111547bc5422fbc6856dca8d140c5b\&vid=41e56e5af5d2f139b8c80632a4c3e859245ee7df\&prid=original\&clid=SAPDAM).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1518), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001518%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/microarray/ancillary-array-equipment/microarray-ancillary-array-equipment-reference_material-list/000001518.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.