> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/rna-enrichment-based-library-prep/library-preparation-rna-enrichment-based-library-prep-faq-list/000003295.md).

# Illumina RNA Prep with Enrichment (L) Tagmentation Sequencing FAQs

The following are commonly asked questions for sequencing Illumina RNA Prep with Enrichment (L) Tagmentation libraries.

**What is the recommended read length and number of reads per sample?**

* Illumina recommends a paired-end run of 76 cycles per read (2x76) with 10 cycles for each index read (Read 1: 76, Index 1: 10, Index 2: 10, Read 2: 76) regardless of the enrichment probe panel or sample input type.
* Libraries can be sequenced with longer read lengths (2x101, 2x126 or 2x151) if a lab would like overlapping reads or additional raw coverage, but it is not required.

**What are the advantages of performing paired-end sequencing of RNA libraries?**\
Paired-end sequencing helps with data alignment, discovery of novel transcripts, and identification of PCR duplicates.

**What are the suggested loading concentrations for the various sequencers?**\
The Illumina RNA Prep with Enrichment [Reference Guide](https://support.illumina.com/downloads/illumina-rna-prep-reference-guide-1000000124435.html) includes select loading concentration recommendations. For other sequencers and panels, consult the denature and dilution guide of the sequencer that will be used.

**Can libraries prepared with this kit be pooled with other library types for sequencing?**\
This is not supported and results are not guaranteed.

**What is the recommended sequencing depth/number of reads per sample?**\
Generally, 3-25 million reads per sample, depending on the panel. The read requirement depends on the enrichment panel used and data needs for each experiment, and may need to be optimized.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3295), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003295%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/rna-enrichment-based-library-prep/library-preparation-rna-enrichment-based-library-prep-faq-list/000003295.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.