> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/rna-enrichment-based-library-prep/library-preparation-rna-enrichment-based-library-prep-faq-list/000003293.md).

# Illumina RNA Prep with Enrichment (L) Tagmentation Input FAQs

**What is the RNA input requirement?**

The protocol is optimized for 10-100 ng of purified total RNA. Lower input amounts can reduce library yield.

**Is the kit compatible with RNA from FFPE samples?**

Yes, for FFPE samples with DV200 value <80% Illumina recommends increasing the number of PCR cycles to 17.

**What are the optimal 260/280 and 260/230 ratios for the input RNA?**

(Refer to the first informational row, the second and third are provided for context.)

<table data-header-hidden><thead><tr><th valign="top"></th><th valign="top"></th><th valign="top"></th><th valign="top"></th></tr></thead><tbody><tr><td valign="top">Substance</td><td valign="top">Absorbance (nm)</td><td valign="top">260/280 Ratio</td><td valign="top">260/230 Ratio</td></tr><tr><td valign="top">Pure RNA</td><td valign="top">280</td><td valign="top">~2.0</td><td valign="top">2.0-2.2</td></tr><tr><td valign="top">EDTA, Carbohydrates, Phenol</td><td valign="top">230</td><td valign="top">&#x3C;1.5</td><td valign="top">&#x3C;2.0</td></tr><tr><td valign="top">Guanidine HCl</td><td valign="top">230</td><td valign="top">&#x3C;1.5</td><td valign="top">&#x3C;2.0</td></tr></tbody></table>

**What is the recommended quantification method for input RNA?**

RNA should be quantified with a fluorometric method specific for RNA.

**Does input RNA need to be treated with DNase?**

Yes, DNA should be removed by including a DNase treatment with the RNA isolation method for sample purity and accurate quantification.

**What is the effect of residual DNA contamination?**

DNA can interfere with accurate quantification of the sample. DNA contamination will also impact the sequencing data, increasing the background level of expression and intergenic reads.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3293), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003293%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/rna-enrichment-based-library-prep/library-preparation-rna-enrichment-based-library-prep-faq-list/000003293.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.