> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/multiomics-library-prep/library-preparation-multiomics-library-prep-faq-list/000009801.md).

# Expected yield and final library traces for SomaSeq Discovery libraries

Note, all information below applies to both SomaSeq Discovery v1 (previously named Illumina Protein Prep) and SomaSeq Discovery v2, unless otherwise indicated. Additionally, the term SomaSeq Automation System will refer to both instrument configurations for the v1 or v2 assay, unless otherwise indicated.

**Why does the library elution volume increase from 100 µl to 200 µl between SomaSeq Discovery v1 and SomaSeq Discovery v2 workflows?**\
This increase in volume is to improve the accuracy of quantification and final library dilution to 1 nM / 2 nM.

**What is the final library QC?**\
There is a mixture of thousands of PCR products representing the SOMAmers captured in the experiment, which should vary in concentration based upon biological variation. Total concentration is measured to ensure the library was created and purified appropriately.

**What are the expected final library yields?**\
User should expect > 2 ng/µl library yield, as stated in the Product Documentation.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9801), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009801%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/multiomics-library-prep/library-preparation-multiomics-library-prep-faq-list/000009801.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.