> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009526.md).

# Illumina Single Cell 3' RNA Prep library QC best practices

**Importance of Library QC**

In the Illumina Single Cell 3' RNA Prep workflow, verifying that both the cDNA and final library pass both QC steps is essential for producing quality sequencing results.

Accurate library quantification is necessary for uniform sample pooling and achieving optimal cluster density during sequencing. Review [Best practices for library quantification](https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000003750.md) to ensure proper QC assessment.

Illumina recommends proceeding with library preparation only after confirming the quality of cDNA from the "Check QC Product" step, when feasible. Refer to [Assessing cDNA QC in Illumina Single Cell 3' RNA Prep](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009524.md) for additional guidance.

**Recommended Library QC Best Practices**

When determining the average library fragment size during the "Check Library" step of the Illumina Single Cell 3' RNA Prep workflow, perform the following for consistent and accurate assessment across samples and experiments.

**1**. **Timing of QC Assessment**

For optimal results, perform quality checks, pooling, and sequencing of purified Illumina Single Cell 3' RNA Prep libraries stored at -25°C to -15°C within 30 days.

If submitting to a core lab for fragment analysis, submit the final libraries with the cDNA QC product.

**2**. **Library Quantification**

Use a Qubit dsDNA High Sensitivity kit to quantify (in ng/µl) 2 µl of each sample.

* **Note**: Do not use the Bioanalyzer or TapeStation to assess concentration.

**3**. **Fragment Analysis**

Use an Agilent Bioanalyzer or TapeStation to determine the average fragment size of the library. Set the region table to 200 bp to 800 bp for library QC analysis.

If necessary, dilute samples to ensure they are within the recommended loading concentrations.

Refer to [Assessing library QC in Illumina Single Cell 3' RNA Prep](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009527.md) to determine if the library is satisfactory to proceed with sequencing.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9526), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009526%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009526.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.