> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009524.md).

# Assessing cDNA QC in Illumina Single Cell 3' RNA Prep

Ensuring that cDNA meets quality expectations during the Check QC Product protocol step is critical for obtaining optimal library QC and sequencing results in Illumina Single Cell 3’ RNA Prep.

Refer to [Illumina Single Cell 3' RNA Prep cDNA QC best practices](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009523.md) to confirm accurate QC assessment.

**cDNA QC Yield Expectations**

* cDNA yield depends on several factors, including sample type, cell/nuclei loading, and PCR cycle number used during the Assess cDNA Quality step.
* Adjust the QC Amplification program to optimize for different input quality and concentration.
* If cDNA QC yields ≥ 200 ng/µl in T10, T20, or T100 kits (not applicable to T2 kits), reduce the number of PCR cycles in the Amplify Library step by 1 or 2 cycles.
* For more information on recommended PCR cycling parameters, refer to [Selecting cycle numbers for High and Low RNA sample types with Illumina Single Cell RNA prep kits](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009521.md)

**cDNA Fragment Analysis Expectations**

The **cDNA average fragment size should be** > **500 bp when the fragment analyzer region table is set from 200 bp to 5000 bp** to proceed with Library Prep.

cDNA peaks, distribution, and average fragment size vary based on sample type and experimental conditions. For example traces from various sample types, refer to [Representative cDNA QC traces for Illumina Single Cell 3' RNA Prep.](https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009525.md)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-0b73584d01f196101cf672b54701136b716bf772%2Fimage1.png?alt=media)

**Figure 1**. Representative human/mouse cell mixture (HEK 293T/NIH 3T3) using a High Sensitivity D5000 ScreenTape.

**If the average cDNA fragment size is** <**500 bp**, **it may indicate significant cDNA degradation**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-2945f456f67684e95cc8fac9ade858d0fdeac095%2Fimage2.png?alt=media)

**Figure 2**. Degraded cDNA QC product from mouse pancreas nuclei that does not pass QC specifications (<500 bp).

Refer to Section 3.2 Quality Checks and Section 3.3 cDNA Degradation in the [Illumina Single Cell 3' RNA Prep Training Packet](https://support.illumina.com/content/dam/illumina-support/courses/illumina-single-cell-3-rna-prep-training-packet/illumina-single-cell-3-rna-prep-training-packet-200065159-02.pdf) on the [Illumina Single Cell Prep Support Page](https://emea.support.illumina.com/sequencing/sequencing_kits/illumina-single-cell-prep.html) (under the Training section) for additional guidance on assessing degraded cDNA.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9524), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009524%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/illumina-single-cell/library-preparation-illumina-single-cell-reference_material-list/000009524.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.